Section 2.0 – Installation
2.1
Unpacking and Inspection
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
Save the box. If there is no apparent damage, unpack the container. Be sure all items shown
on the packing list are present. If items are missing, notify Rosemount immediately.
2.2
Installation
2.2.1
General Information
1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in
areas of extreme temperatures is not recommended unless a sunshield is used. Make
sure to note the Ambient temperature specifications in section 1. The analyzer cannot
be operated in ambient (shaded) conditions greater than 60 °C.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency
interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be
sure there is easy access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. See Figures 2-1 and 2-2.
5. Install cable gland fittings and plugs as needed to properly seal the analyzer on all six en-
closure openings. The USB port cover must be fully installed on the front cover to ensure
proper analyzer sealing.
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol
appears, consult the manual for appropriate action.
WARNING
Installation
15
Instruction Manual
Section 2: Installation
LIQ-MAN-56
April 2017
Summary of Contents for Rosemount 56
Page 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Page 2: ......
Page 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Page 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Page 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Page 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Page 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Page 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Page 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...