Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
3-21
Previous Constant
Refer to Figure 3-27. Use this screen to examine the cell zero constant stored
from the second to last calibration check. The constant represents the voltage
generated by the cell when no difference exists between the amount of O
2
on
the reference and process sides of the cell. Press NEXT to access the MAX
CELL T screen.
Max Cell T
Refer to Figure 3-28. Use this screen to examine the maximum temperature
attained by the O
2
cell. This value can be reset under the PROGRAM MENU.
Pressing NEXT accesses the SW VER screen.
Figure 3-27. Previous Constant
Figure 3-28. Max Cell T