Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
3-20
Current Constant
Refer to Figure 3-25. Use this screen to examine the cell zero constant
calculated from the most recent calibration check. The constant represents
the voltage generated by the cell when no difference exists between the
amount of O
2
on the reference and process sides of the cell. Press NEXT to
access the PREVIOUS SLOPE screen.
Previous Slope
Refer to Figure 3-26. Use this screen to examine the slope value stored from
the second to last calibration check. The slope is the amount of cell voltage
generated for a given O
2
value. For each calibration check, record the slope
over the life of the probe. Tracking the slope will indicate if the probe is
degrading. Press NEXT to access the PREVIOUS CONSTANT screen.
Figure 3-25. Current Constant
Figure 3-26. Previous Slope
Summary of Contents for Rosemount 5081FG
Page 2: ......
Page 6: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG TOC 4 ...
Page 20: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 1 10 ...
Page 36: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 16 ...
Page 53: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 17 Figure 3 20 Show Fault ...
Page 82: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 6 ...
Page 94: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG A 8 ...
Page 96: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 98: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Page 100: ......
Page 103: ...Index 2 ...