Troubleshooting
5-7
HART Error Messages
Description
Hart Application Error... “hc.ddl device revision x
not found.”...Refresh System Card...See User's
Manual and Programming Interface for details.
An expected version of hc.ddl is not found.
Hart Application Error... “registry failure”...Reinstall
System Files...See User's Manual for details.
The information describing the location of the
DDs is not found in the registry.
Voltage is currently detected at the Fieldbus
Terminals. Reconfigure unit and try again.
The HART Application finds voltage on the
fieldbus connector during initialization and
during reconnection to a device.
CDC Error...“failure while switching to
HART”...Reinstall System Files...See User’s
Manual for details.
The HART Application fails to put the CDC in
HART mode for talking to a HART device.
DD Error…"tokenizer mismatch"…DD version
<manufacture name, device name, dev. rev. #, and
DD rev. #>
; HART application
<version x.y>
; HART
app version x.y...Reinstall DD…See Programming
Interface for details.
The version of the tokenizer used to tokenize
the DD is not what the HART Application
expects.
DD Error…"linker mismatch"...DD version x.y;
HART app version x.y...Reinstall DD…See
Programming Interface for details.
The version of the linker used to link the DD
is not what the HART Application expects.
DD Error…"developed-to mismatch"…DD version
<manufacture name, device name, dev. rev. #, and
DD rev. #>
; HART application
<version x.y>
…See
Programming Interface for solution.
The DD was developed and tested using a
later HART Application revision.
Warning: this DD is untested with the current Hart
App ver
<manufacture name, device name, dev.
rev. #, and
DD rev. #>
…you may experience unexpected or
inappropriate results…See Programming Interface
for details on DD updates…Do you want to accept
these risks and proceed?
The DD was not tested and verified with the
HART Application.
DD Error…"tested-to sequence mismatch"…DD
version
<manufacture name, device name, dev.
rev. #, and
DD rev. #>
; HART application
<version x.y>
; HART
app version x.y…See Programming Interface for
solution.
The DD was tested with a previous version,
but not with this version of the HART
Application.
DD Error…"developed-to mismatch"...DD version
<manufacture name, device name, dev. rev. #, and
DD rev. #>
; HART application
<version x.y>
; HART
app version x.y...Reinstall DD…See Programming
Interface for details
The DD was developed using a later HART
Application revision.
DD Error…"DD header file corrupted"...Reinstall
Device version
<manufacture name, device name,
dev. rev. #, and
DD rev. #>
; HART application
<version x.y>
…See
Programming Interface for details.
There is a check sum error in the DD header
file.
Hart Application Error…"data compression dll is
missing"…Reinstall System Files...See User's
Manual for details.
The data compression component cannot be
located.
Summary of Contents for Rosemount 375
Page 1: ...USER S MANUAL 375 Field Communicator User s Manual ...
Page 2: ......
Page 4: ......
Page 10: ...Introduction 1 2 ...
Page 40: ...Learning the Basics 2 30 ...
Page 112: ...Product Certifications B 8 ...
Page 113: ...Product Certifications B 9 ...
Page 114: ...Product Certifications B 10 ...
Page 132: ...G viii ...