Fieldbus Functionality
4-18
Display Network Settings
To view the network parameters for the device or
segment:
1. Display the Live Device List; see page 4-9.
2. Double-tap a device.
3. Double-tap
Advanced
.
4. Double-tap
Network Management
.
5. Double-tap the desired label to view the values.
Tap the
SAVE
button to save this as a text file. Transfer
this file to a PC using the Easy Upgrade Programming
Utility. The file appears in the User Data tab.
Change the I/O Block Schedule
NOTE
You cannot schedule I/O blocks when, at a minimum,
any of the following hosts/linking devices are detected
on the segment: Rosemount 3420, DeltaV, Ovation,
Allen-Bradley, Honeywell Experion PKS, or Yokogawa
Stardom. To schedule I/O blocks, use the host system or
remove the device from the segment and connect it
directly to the 375 Field Communicator.
To change the schedule:
1. Display the Live Device List; see page 4-9.
2. Double-tap a device.
3. Double-tap
Advanced
.
4. Double-tap
Schedule
. The I/O Block Schedule
window appears. Changing the I/O Block Schedule
would typically be done on the bench (not connected
to a control system) when you want to execute an I/O
block to verify that the outputs of the block are being
computed correctly.
5. Using the drop-down menu, select the appropriate
MacroCycle.
6. Tap the checkbox next to the I/O blocks that are to be
scheduled.
7. Tap
OK
. This will schedule the blocks and set the
mode on these blocks to Auto.
8. After reading the note, tap
OK
to acknowledge that
the I/O blocks were successfully scheduled.
Summary of Contents for Rosemount 375
Page 1: ...USER S MANUAL 375 Field Communicator User s Manual ...
Page 2: ......
Page 4: ......
Page 10: ...Introduction 1 2 ...
Page 40: ...Learning the Basics 2 30 ...
Page 112: ...Product Certifications B 8 ...
Page 113: ...Product Certifications B 9 ...
Page 114: ...Product Certifications B 10 ...
Page 132: ...G viii ...