Set simulation duration
The simulation duration parameter sets the length in time of the simulation.
Procedure
1. Select
S.Tim
from the menu.
2. Set time span for simulation.
Setting range: 1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60 minutes
Factory setting: three minutes
3. Press
Enter
.
Switch simulation on or off
Switching the simulation on starts the simulation and switching it off cancels the
simulation.
Procedure
1. Select
S.On
from the menu.
2. Select
On
or
OFF
.
3. Press
Enter
.
Note
The simulation is active until
Enter
is pressed again or the time set via
S.Tim
elapses. During the simulation, the level transmitter displays
SIM
every three
seconds. After the simulation, the level transmitter goes again to the parameter
S.On
and internally the level transmitter goes again to the operating mode (and the
process value transmission).
After another 30 seconds the display goes again to the process value display.
Note
If using IO-Link to start and configure a simulation, it can be canceled only using IO-
Link. The level transmitter displays
C.Loc
if the simulation is attempted to be
canceled with the function buttons.
3.5
Using the transmitter
The procedures in this section provide instructions for operating the level transmitter.
3.5.1
Operation with a single probe
Using the level transmitter with a single probe is suited for the detection of aqueous
media, especially heavily soiled aqueous media.
Note
The level transmitter is intended only for operation with a single probe. A coaxial probe is
not available for this unit.
Operation
Quick Start Guide
January 2019
00825-0100-4326
38
Quick Start Guide
Summary of Contents for Rosemount 326L
Page 1: ...Quick Start Guide 00825 0100 4326 Rev 01 January 2019 Rosemount 326L Level Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4326 4 Quick Start Guide ...
Page 12: ...Introduction Quick Start Guide January 2019 00825 0100 4326 12 Quick Start Guide ...
Page 24: ...Installation Quick Start Guide January 2019 00825 0100 4326 24 Quick Start Guide ...
Page 42: ...Operation Quick Start Guide January 2019 00825 0100 4326 42 Quick Start Guide ...
Page 48: ...Setting ranges Quick Start Guide January 2019 00825 0100 4326 48 Quick Start Guide ...
Page 50: ...Product certifications Quick Start Guide January 2019 00825 0100 4326 50 Quick Start Guide ...