2. Mark the desired length (L) on the probe. The reference point is the lower edge of
the process connection.
3. Remove the probe from the unit. Do not lose the O-ring between the probe
attachment piece and the probe.
4. Shorten the probe at the mark.
5. Remove all burrs and sharp edges. For hygienic requirements: Restore the required
surface quality. If necessary, polish the probe.
Note
In cases with strong vibrations or moving viscous medium, it may be necessary to
secure the probe connection using screw retaining compound. In this case, ensure
that the compound is harmless because it might migrate into the medium.
Postrequisites
Reinstall the probe by following the steps in
2.2.2
Install the probe
Install the probe onto the level transmitter to ensure accurate level readings.
For more information, see
.
Note
The probe is not included with the level transmitter.
CAUTION
Do not damage the surfaces of the process connection and probe. Use tools suitable for
use with plastic surfaces.
Procedure
1. Remove the protective cover and protective devices from the unit and the probe.
Installation
Quick Start Guide
January 2019
00825-0100-4326
18
Quick Start Guide
Summary of Contents for Rosemount 326L
Page 1: ...Quick Start Guide 00825 0100 4326 Rev 01 January 2019 Rosemount 326L Level Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4326 4 Quick Start Guide ...
Page 12: ...Introduction Quick Start Guide January 2019 00825 0100 4326 12 Quick Start Guide ...
Page 24: ...Installation Quick Start Guide January 2019 00825 0100 4326 24 Quick Start Guide ...
Page 42: ...Operation Quick Start Guide January 2019 00825 0100 4326 42 Quick Start Guide ...
Page 48: ...Setting ranges Quick Start Guide January 2019 00825 0100 4326 48 Quick Start Guide ...
Page 50: ...Product certifications Quick Start Guide January 2019 00825 0100 4326 50 Quick Start Guide ...