The display shows the initial display.
2. Select
LEnG
from the menu.
If the parameter does not contain a value, the display shows
nonE
.
3. Press and hold up or down for at least one second. The display shows the detected
probe length.
4. If the detected probe length is incorrect, press up or down to change the probe
length to the correct value.
5. Press
Enter
.
The level transmitter proceeds to operating mode.
Notes
The level transmitter can automatically detect the length of the probe if:
•
The tank is empty
•
The level transmitter is mounted on a sufficiently large launching plate
For instructions on manually determining the length of the probe, see
3.4.2
Perform a tank adjustment
Tank adjustment reduces the effect of interference and ensures a higher excess gain in
difficult application conditions.
For more information on tank adjustments, see
.
Prerequisites
The tank adjustment menu option
tREF
appears only if the set probe length is greater than
or equal to 10.24 in (260 mm).
Procedure
1. Select
tREF
from the menu.
The display shows
nonE
or the value set by a previous tank adjustment.
2. Press and hold up or down for at least one second.
The display shows the distance value. The default value is 0.39 in (10 mm).
3. If necessary, correct the value using the up and down buttons.
4. Press
Enter
.
The display shows
donE
.
5. Press
Enter
again.
The level transmitter reboots and returns to operating mode.
3.4.3
Configure the display
Modify how the level transmitter displays information.
Procedure
1. Change the unit of measurement.
Quick Start Guide
Operation
00825-0100-4326
January 2019
Quick Start Guide
31
Summary of Contents for Rosemount 326L
Page 1: ...Quick Start Guide 00825 0100 4326 Rev 01 January 2019 Rosemount 326L Level Transmitter ...
Page 2: ...2 ...
Page 4: ...Contents Quick Start Guide January 2019 00825 0100 4326 4 Quick Start Guide ...
Page 12: ...Introduction Quick Start Guide January 2019 00825 0100 4326 12 Quick Start Guide ...
Page 24: ...Installation Quick Start Guide January 2019 00825 0100 4326 24 Quick Start Guide ...
Page 42: ...Operation Quick Start Guide January 2019 00825 0100 4326 42 Quick Start Guide ...
Page 48: ...Setting ranges Quick Start Guide January 2019 00825 0100 4326 48 Quick Start Guide ...
Page 50: ...Product certifications Quick Start Guide January 2019 00825 0100 4326 50 Quick Start Guide ...