52
Reference Manual
00809-0700-4001, Rev AA
Operation and Maintenance
May 2017
Operation and Maintenance
Figure 5-5. Analog Output Trim Example
A. Before trim
B. After Trim
5.6.1
Performing digital-to-analog trim (4–20mA output trim)
Note
If a resistor is added to the loop, ensure that the power supply is sufficient to power the transmitter to a
20 mA output with additional loop resistance. Refer to
Performing a 4–20 mA output trim with a Field Communicator
From the
HOME
screen, enter the Fast Key sequence and follow the steps within the Field Communicator
to complete the 4–20 mA output trim.
Performing a 4–20 mA output trim with AMS Device Manager
Right click the device and, under the
Method
drop down menu, move cursor over
Calibrate
and select
Analog Calibration
.
1. Select
Digital to Analog Trim.
2. Follow the screen prompts to perform a 4–20 mA output trim.