Instruction Manual
IM-106-880, Rev 1.0
January 2007
8-5
OCX 8800
SB TC Open
(Sample block heater thermocouple open)
Probable Cause
Recommended Corrective Action
Sample block thermocouple or thermocouple
circuit open
Check sample block thermocouple and circuit wires for breaks or loose connections per
Figure 8-3 and Figure 7-4. Repair breaks or loose connections or replace failed
thermocouple. Perform Reset procedure in Section 3, Configuration and Startup to
continue operation.
O2 TC Shorted
(Oxygen sensor heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
O
2
thermocouple or thermocouple circuit shorted
Check O
2
thermocouple and circuit wires for short circuit condition per Figure 8-3 and
Figure 7-4. Repair shorted wiring or replace failed thermocouple. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
Slow heatup during cold start
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the O2 Htr Failure alarm procedure.
COe TC Shorted
(Combustibles sensor heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
Combustibles sensor thermocouple shorted
Check thermocouple resistance (lead to ground) of reference and active thermocouples
per Figure 8-3 and Figure 7-5. If either thermocouple is open or shorted to ground,
replace combustibles sensor.
Slow heatup during cold start
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the COe Htr Failure alarm procedure.
SB TC Shorted
(Sample block heater thermocouple shorted)
Probable Cause
Recommended Corrective Action
Sample block thermocouple or thermocouple
circuit shorted
Check sample block thermocouple and circuit wires for short circuit condition per
Figure 8-3 and Figure 7-4. Repair shorted wiring or replace failed thermocouple.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
Slow heatup during cold start
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
If alarm persists, refer to the SB Htr Failure alarm procedure.
O2 TC Reversed
(Oxygen sensor heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
O
2
thermocouple wires reversed
Check O
2
thermocouple wiring per Figure 8-3 and Figure 7-4. Check the wiring at the
sensor and inside the electronics housing. Correct reversed-wires fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.
COe TC Reversed
(Combustibles sensor block heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
Combustibles thermocouple wires reversed
Check combustibles thermocouple wiring per Figure 8-3 and Figure 7-5. Check the
wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
SB TC Reversed
(Sample block heater thermocouple reversed)
Probable Cause
Recommended Corrective Action
Sample block thermocouple wires reversed
Check sample block thermocouple wiring per Figure 8-3 and Figure 7-4. Check the
wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.
Table continued on next page
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Page 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Page 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Page 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Page 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Page 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...