Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
3-4
SET TEST GAS VALUES
Use HART/AMS or the optional LOI to set test gas values for calibration.
Refer to Section 4, Using the LOI or Section 5, Using HART Communications
for more information.
Setting Test Gas Values with HART
1. Use the HART communicator or AMS software to access the HART
menu.
2. From the
DETAILED SETUP
menu, select
O2 CALIB PARAMS
.
3. From
O2 CALIB PARAMS
, select
O2 High Gas
. Enter the percent O
2
used for the high O
2
test gas.
4. From
O2 CALIB PARAMS
, select
Low TG
. Enter the percent O
2
used
for the low O
2
test gas.
5. From the
DETAILED SETUP
menu, select
COe CALIB PARAMS
.
6. From
COe CALIB PARAMS
, select
COe Test Gas
. Enter the CO
concentration (ppm) used for COe test gas.
Setting Test Gas Values with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the
SYSTEM
menu, select
Calib Setup
.
3. From
Calib Setup
, select
O2 High Gas %
. Enter the percent O
2
used
for the high O
2
test gas.
4. Select the down arrow and the next selection will be
O2 Low Gas %
.
Enter the percent O
2
used for the low O
2
test gas.
5. Select the down arrow several times to display
COe Test Gas
. Enter the
CO concentration (ppm) used for COe test gas.
OCX 8800 RESET
PROCEDURE
Whenever you correct an equipment alarm or fault condition, the OCX 8800
will either revert to normal operation or continue to indicate an alarm status
condition. If the equipment does not revert to normal operation when a fault
condition is cleared, or if instructed to do so in Section 8, Troubleshooting,
use the following procedure to reset the OCX 8800.
OCX Reset with the LOI
1. Use the "Z" pattern to enter the LOI menu tree. (Refer to Section 4,
Using the LOI).
2. Select the
SYSTEM
submenu.
3. From the
SYSTEM
submenu, select the
Status
submenu.
4. From the
Status
submenu, select
Reset Device
. The OCX 8800 will
reset and the LOI will revert to the normal operation display.
OCX Reset with HART
Remove the OCX 8800 from the process loop and recycle power.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Page 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Page 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Page 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Page 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Page 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...