Instruction Manual
IM-106-880, Rev 1.0
January 2007
7-23
OCX 8800
Figure 7-19. O
2
Cell,
Thermocouple, and
Heater Connections
2. If thermal switch (15) was removed, install thermal switch on mating
heater clamp (12), connect wires of heater rods (13), and install
insulator (14).
3. Install the heater rods (13), heater clamps (12), and screws (11).
4. Reconnect the heater rod leads at the sensor housing terminal blocks,
(Figure 7-19).
+
+
+
+
O2
Thermocouple Wires
ORG
RED
T/C O2
T/C CO
T/C SB
EXC
CJC
O Cell Wires
2
O Heater Wires
2
Return
Wire
O Cell and Heater Strut
Assembly
2
CO REF
CO ACT
Sensor Housing
T
erminals
+
-
HTR 02
HTR SB
2
GRN
2
1
+
-
+
-
+
-
+
-
+
-
+
YEL
-
+
+
37390069
HTR 02
2
1
Sample Block
Heater Rods
Thermal
Switch
HTR CO
1
EXC
-
Sample
Block
Thermo-
couple
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Page 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Page 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Page 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Page 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Page 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...