Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
6-6
5. From the
SELECT ACTION
screen, select
START/NEXT CALSTEP
to
continue calibration, select
ABORT CAL
to abort calibration or
EXIT
CAL
to exit calibration. Select one item from the list and press
ENTER
.
OCX: TAG NAME
SELECT ACTION
1. START/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
6. When the Calibration Status is at the
AppO2Low
step, switch on O
2
Low Gas. Verify the O
2
concentration measured matches the
O2 LOW
GAS
parameter in the Setup. Press
OK
when ready.
7. Select
Start/Next Cal Step
to start applying the O
2
Low Gas. The time
to apply the test gas is specified by the
Gas Time
.
8. The Calibration Status should be automatically changed to
FlowO2Low
and then
ReadO2Low
for a period of time. During this period, if an
attempt is made to go to the next calibration step by pressing
OK
and
selecting
Start/Next Cal Step
, you will be prompted with
Operator step
command is not accepted at this time
. The
Next Cal Step
command
is not accepted at this time.
9. When ready, Calibration Status will stop at the
AppO2Hi
. Switch off the
O
2
Low Gas and switch on the O
2
High Gas. Verify the O
2
concentration
measured matches the
O2 HIGH GAS
parameter in the Setup. Press
OK
when ready.
10. Select
Start/Next Cal Step
to start applying the O
2
High Gas. The time
to apply the test gas is specified by the
Gas Time
.
11. The Calibration Status should be automatically changed to
FlowO2Hi
and then
ReadO2Hi
for a period of time. During this period, if an attempt
is made to go the next calibration step by pressing
OK
and selecting
Start/Next Cal Step
, you will be prompted with
Operator step
command is not accepted at this time
. The
Next Cal Step
command
is not accepted at this time.
12. When ready, Calibration Status will stop at
STOP GAS
. Switch off the
O
2
High Gas. Press
OK
when ready. Select
Start/Next Cal Step
to start
purging gas. The time to purge gas is specified by the
Purge Time
.
13. When the
Purge
step is complete, the Calibration Status will be at
IDLE
if the calibration is successful or
CAL RECOMMENDED
if the
calibration has failed. A Calibration Failed alarm will be set if the
calibration has failed.
14. When calibration is complete. Select
Exit Cal
to exit the calibration
method.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 50: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 4 10 ...
Page 70: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 6 10 ...
Page 114: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 8 12 ...
Page 124: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 9 10 ...
Page 156: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 A 32 ...
Page 164: ...Instruction Manual IM 106 880 Rev 1 0 January 2007 OCX 8800 C 2 ...