32
Micro Motion
®
Model D and DT Sensors Instruction Manual
Wiring
continued
Model D600 wiring to RFT9712 transmitter
Model D600 sensor wiring to the 9-wire Model 1700 or 2700 transmitter
Core processor to a 4-wire
remote transmitter or remote
host
To connect wiring at the core processor:
1. Use one of the following methods to shield the wiring from the core
processor to the remote transmitter:
• If you are installing unshielded wiring in continuous metallic conduit
that provides 360° termination shielding for the enclosed wiring, go
to Step 6, page 35.
• If you are installing user-supplied cable gland with shielded cable or
armored cable, terminate the shields in the cable gland. Terminate
both the armored braid and the shield drain wires in the cable gland.
• If you are installing a Micro Motion-supplied cable gland at the core
processor housing:
-
Prepare the cable and apply shielded heat shrink as described
below. The shielded heat shrink provides a shield termination
suitable for use in the gland when using cable whose shield
consists of foil and not a braid. Proceed to Step 2.
9
7 5
3
1
8
6 4 2 0
9
7 5
3
1
8
6 4 2 0
9
7 5
3
1
8
6 4 2 0
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Model D600 sensor
terminals
Flowmeter
cable
RFT9712
terminals
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Brown
Orange
Green
Violet
White
Gray
Blue
Yellow
Red
Black (Drains)
Maximum cable length 1000 ft. (300 m)
Prepare cable in accordance with the instructions
that are shipped with the cable
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Model D600 sensor
junction box
terminals
Flowmeter
cable
Field-mount Model
2700 terminals
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Brown
Violet
Yellow
Maximum cable length 60 ft. (20 m)
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Blue
Gray
Orange
Red
Green
White
Ground screw
Black
Summary of Contents for Micro Motion
Page 12: ...8 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 16: ...12 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 42: ...38 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 62: ...58 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 64: ...60 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 68: ...64 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 70: ...66 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 73: ......