30
Micro Motion
®
Model D and DT Sensors Instruction Manual
Wiring
continued
1. Locate the wires by color and terminal number.
2. Insert the stripped ends of the individual wires into the terminal
blocks. No bare wires should remain exposed.
• At the sensor, connect wiring inside the junction box.
• At the transmitter, connect wiring to the transmitter’s intrinsically
safe terminals for sensor wiring.
3. Tighten the screws to hold the wires in place.
4. Ensure integrity of gaskets, then tightly close and seal the
junction-box cover and all housing covers on the transmitter.
Model D600 wiring to Model 3500 with I/O cable
Model D600 wiring to Model 3500 with screw or solder terminals
CAUTION
Drain wires from a 9-wire cable must be clipped at the
sensor end and insulated with heat-shrink wrapping.
Failure to properly terminate drain wires will cause
sensor error.
Br
o
w
n
Red
Or
a
nge
Y
e
llo
w
Gre
e
n
Bl
ue
Vi
ole
t
Gr
a
y
Wh
it
e
B
la
c
k (drain
s)
Connect outer
braid of shielded
or armored cable
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Model D600 sensor
junction box
terminals
Model 3500
with I/O cable
Not approved for intrinsic safety
in Europe
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Model 3500
with screw-type or
solder-tail terminals
Yellow
Violet
Green
Blue
Brown
Black (Drains)
Orange
White
Gray
Red
c4
c6
c8
c10
c12
a4
a6
a8
a10
a12
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Model D600 sensor
junction box
terminals
Prepare cable in accordance with the
instructions that are shipped with the cable.
Do not allow shields to contact sensor
junction box.
Summary of Contents for Micro Motion
Page 12: ...8 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 16: ...12 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 42: ...38 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 62: ...58 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 64: ...60 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 68: ...64 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 70: ...66 Micro Motion Model D and DT Sensors Instruction Manual ...
Page 73: ......