100-412-183 REV. 13
109
3
Go to the P/S Diagnostics tab. Press the Start Test button to start the test.
Sonics will become active and the button will change to Stop Test. Press the
button again to stop the test.
If the OK - Memory Stored indicator becomes active the test procedure is
finished.
NOTICE
Low level ultrasonic energy will remain active until the Stop Test button is
pressed. If you close the web browser, ultrasonic energy will turn off
automatically after 5 seconds.
4
If the Overload - Memory Cleared indicator becomes active the test will be
interrupted, press the Reset Overload button and repeat step 3 one time
only. If the alarm persists, refer to
.
Table 7.12 Power Supply Ultrasonic Test Procedure (Web Page Interface)
Step
Action
Summary of Contents for DCX S
Page 4: ...iv 100 412 183 REV 13 ...
Page 22: ...12 100 412 183 REV 13 ...
Page 36: ...26 100 412 183 REV 13 ...
Page 47: ...100 412 183 REV 13 37 4 3 Declaration of Conformity Figure 4 1 Declaration of Conformity ...
Page 50: ...40 100 412 183 REV 13 ...
Page 54: ...44 100 412 183 REV 13 Figure 5 1 DCX S Power Supply Benchtop Dimensional Drawing ...
Page 86: ...76 100 412 183 REV 13 ...
Page 87: ...100 412 183 REV 13 77 Chapter 6 Converters and Boosters 6 1 Converters and Boosters 78 ...
Page 120: ...110 100 412 183 REV 13 ...
Page 137: ...100 412 183 REV 13 127 8 5 Circuit Diagram Figure 8 2 Interconnect Diagram Power Supply ...
Page 144: ...134 100 412 183 REV 13 ...
Page 145: ...100 412 183 REV 13 135 Appendix A Timing Diagrams A 1 Timing Diagrams 136 ...
Page 147: ...100 412 183 REV 13 137 Appendix B Signal Diagrams B 1 Signal Diagrams 138 ...
Page 154: ...144 100 412 183 REV 13 ...