1B-12 / DP Transmitters
2808B
An on-site, zero test signal for calibration purposes can be generated by opening the bypass
valve and closing both shutoff valves. Calibration information will be found in Section 3.
Figure 1B-9 - Gas, Horiz, Pipe
Figure 1B-10 - Gas, Vert. Pipe
Steam Application
. When measuring steam pressure, the maximum temperature of the
Transmitter's electronic circuitry must be strictly observed. Temperatures above the
specified limit (see topic 2.1, Environmental Temperature) will cause output errors and
possibly result in damage to the Transmitter. One method of protecting the Transmitter
can be achieved by installing an extended, liquid-filled connecting line as shown in Figures
1B-11 and 1B-12. The liquid functions as a buffer and prevents live steam from entering the
transmitter.
When using liquid-filled system, the connecting line must be installed in a descending step
so that the Transmitter is below the level of the process pipe tap and filling tee; this slope
will maintain the liquid in the connecting line and prevent it from being drawn into the
process pipe. Liquid-filled lines must also be properly filled and bled, and checked on a
regular basis.
Figure 1B-11 - Steam Process, Horiz. Pipe Fig. 1B-12 - Steam Process, Vert. Pipe
Summary of Contents for Bristol 2808 Series
Page 12: ...BLANK PAGE ...
Page 14: ...1 2 Introduction 2808B Figure 1 1 2808 XXB Transmitters ...
Page 16: ...BLANK PAGE ...
Page 30: ...BLANK PAGE ...
Page 32: ...1B 2 DP Transmitters 2808B Figure 1B 1 Model 2808 35B Figure 1B 2 Transmitter Assemblies ...
Page 40: ...1B 10 DP Transmitters 2808B Figure 1B 6 Transmitter Housing Rotation Diagram ...
Page 54: ...BLANK PAGE ...
Page 57: ...2808B Calibration 3 3 Figure 3 2 Calibration Adjustments ...
Page 60: ...3 6 Calibration 2808B TABLE 3 A COARSE ZERO SWITCH SETTINGS Elevation ...
Page 61: ...2808B Calibration 3 7 TABLE 3 B COARSE ZERO SWITCH SETTINGS Suppression ...
Page 66: ...BLANK PAGE ...
Page 68: ...4 2 Service 2808B Figure 4 1 Using Internal TEST Terminals to Measure Current ...
Page 72: ...BLANK PAGE ...
Page 74: ...BLANK PAGE ...
Page 90: ...BLANK PAGE ...
Page 92: ...BLANK PAGE ...
Page 100: ...BLANK PAGE ...
Page 128: ...BLANK PAGE ...
Page 139: ...BLANK PAGE ...