Service & Clean
Enter this menu by selecting “MENU’” from the main screen, enter the PASSCODE and select “SERVICE” or “CLEAN”.
Save the settings by selecting “ESC”, then confirm the changes by moving to “YES” and pressing the knob.
Service
From this menu it is possible to view the functions of the inputs relative to:
pH or ORP Probe (as specified when ordering)
Cl2 Probe
Temperature Probe
Instrument’s MAC address and 6 DIGITS Code for ERMES Communication Service
Clean
From this menu you can start the cleaning process of the reaction tank by filling it and emptying it with water.
1. Select this menu to start the procedure
2. The solenoid valve (9) to load water to the reaction tank "EV H
2
O
2
: ON” is active
3. The systems displays the message "PLEASE OPEN DRAIN". Open drain tap (11).
4. When the reaction tank is full, the water loading solenoid valve (9) stops "EV H
2
O
2
: OFF”
5. The solenoid valve (7) of the reaction tank and the instrument displays "EV BLEED: ON”
4. Open the drain tap (11) on the storage tank (10) to drain the water
5. The procedure ends and the instrument displays "EV BLEED: OFF”
Priming
Enter into this menu to manually enable or disable pumps and outputs. This function can be used for initial pumps priming.
At the end of the procedure it is recommended to empty both the tanks: storage and reaction.
Note: if the storage tank isn’t completely empty, the EV for
chlorine dioxide
discharging will not turn on.
Rotate knob, highlight output (Relay) or pump (Out) and press knob to change its own status.