EMC INDUSTRIAL GROUP LTD
Inputs
MW61A_IM_ALL_SV6.09d_en
25/47
Weight change constants
Q2181 control function
8088, g0
4
0
The control functions sets whether the flowrate is measured by weight loss (decreasing weight) or
weight gain (increasing weight).
Set to 0 for weight loss or to 1 for weight gain.
Clock
The unit has a real time clock. The clock is used to date stamp printouts and can cause printouts (or
other events) to occur at user set times during the day or week. The clock will run continue to run for a
few days or weeks without power. The time is also shown in the top right corner of the display (if no
alarms are present).
Q2191 clock enable
9910
0
Set to 0 to disable the clock or 1 to enable the clock. When disabled, time & date information is not
shown on any printouts.
Daylight saving
Q21921 locality
9912
0
Selects the locality for daylight saving correction. Use the UP & DOWN keys to select a locality or set to
0 to disable daylight saving correction or set to CUSTOM to set your own correction dates. If CUSTOM is
selected, the following three settings specify when the correction will be made.
Q21922 weekday & time for DST
9914
5
Selects the day of the week and time when the daylight saving correction will be made.
Q21923 start of daylight saving
9916
0
Selects the time of the year when daylight saving will start.
Q21924 end of daylight saving
9918
0
Selects the time of the year when daylight saving will end.
Q21925 daylight saving time (DST)
9920
0
Shows if daylight saving is on or off. If automatic daylight saving correction is off, then this setting may
be used to move the clock forwards or backwards by one hour.
Set clock
Q21931 set date
9922
Sets the date. For example to set 9 March 2015, enter 150309.
Q21932 set time
9924
Sets the time in 24hour format.
Inputs
Basic Settings
Inputs
Internal Signals
Outputs
Communications & Display
Information, Resets & Final Calibration
Digital inputs
The following settings select the function of the digital inputs. The hint line shows (off) when there is
no voltage present, and (on) when voltage is applied to the input.
The UP and DOWN keys can be used to select the function from the available list.
The list is shown in the
Instruction Manual
in the ‘I/O Function Table’.
Q2211 IN1 function
(& state)
6
8170, g3
7
acquire zero=16
Q2212 IN2 function (& state)
8172, g3
acquire tare=17
Q2213 IN3 function (& state)
8174, g3
print=40
Q2214 IN4 function (& state)
8176, g3
capture weight=47
Q2215 IN5 function (& state)
8178, g3
print total=41
Q2216 IN6 function (& state)
8180, g3
reset total=32
Q2217 IN7 function (& state)
8182, g3
totalise=33
Q2218 IN8 function (& state)
8184, g3
hold flowrate=8
Q2219 IN9 function (& state)
8186, g3
acquire zero=16
Q2210 IN0 function (& state)
8190, g3
no function=0
4
0=weight loss, 1=weight gain
5
Defaults to whatever the current setting is.
6
Hint line shows state as (on) or (off)
7
See I/O Function Table page 40. Use negative numbers to reverse the signal sense eg 1=stop (stop when input is on), -1=NOT stop
(stop when input is off).