15.1.1 – MAXIMUM ALARM SETTING
Display shows
MAX
and current measured value both flashing.
Press – or + to select max measured value ALARM point
Press OK to confirm
15.1.2 – MINIMUM ALARM SETTING
Display shows
min
and current measured value both flashing.
Press – or + to select minmeasured value ALARM point
Press OK to confirm
15.1.3 – ALARM SETTING
Display shows
OVER DELAY and time counter h:min flashes.
Press – or + to select Over Dosing time, passed this point
pump will enter alarm mode and stop dosing operations.
Press OK to confirm
Press ESC to return measuring mode
16.0 – RESET / STAND-BY
16.1 – RESET PROCEDURE
Here following are steps for RESET procedure:
• Switch OFF and ON the unit
Press at the same time ESC and OK until display will show
the latest firmware revision (e.g.02.3.3), then will appear
SET UP and default parameters will blink.
Now user has 15 seconds to complete reset operation.
press together
After 15 seconds from entering SETPOINT menu, it’ NOT possible to activate reset procedure. Therefore the
user has to switch unit OFF and then ON again to start reset procedure.
16.2 – STAND-BY PROCEDURE
The unit can be placed in STAND-BY mode to allow sensors and
accessories maintenance operations.
Press and hold + and – at the same time until display will
show STAND-BY STBY
flashing
• To exit Stand-By mode
Press and hold + and – buttons at the same time until
display will show previous measuring settings.
press together
TO EXIT
press together