10.0 – PROGRAMMING CONTROLLER
KEEP IN MIND:
if no selection is made dusring set-up step, pump wil start to release pulse strokes
10.1 – PUMP SET UP / CONTROLLER SETTING
The first operation is to select the required parameter intended to measure and control pH or RX (mV).
Pump power ON, display will show last software revision and also the
type of measurement parameter flashing (default setting: pH with
operating menu BASE basic-simplified mode).
Now it is possible to select required measuring parameter, pH or
RX, by means of – or + buttons and press OK to confirm.
Select type of measuring parameter and confirm OK, programming
moves to measuring mode, display will show current measured value.
Press OK to enter menus; it will show SETPOINT menu,
using – or + buttons to scroll all 4 menus.
Set point menu
Calibration menu
Alarm menu
Set up menu
- this symbol “pointing finger” indicates programming steps “PRESS”.
NOTE
After about 2 seconds, selection is operational and operator can decide whether to quit SETUP menu or to
continue for additional settings. In view of selecting pH, Redox measuring, check next paragraph.
To enter SET UP menu, display will flash SETUP
scroll with + or – buttons and once reached required menu
press OK to confirm
After about 2 seconds, selection is operational and operator can
decide whether to quit SETUP menu or to continue for additional
settings. In view of selecting pH, Redox measuring, check next
paragraph.