4.
Operation
4.1.
Description of the menus
The simulator is operated by an easy menu control system. Seven function
keys are available to select parameters and functions.
The selected parameter is blinking and can be changed by turning the knob (incr./decr.). The digit
to be changed can be selected with the cursor (
çè
).
ESC will take you back to the previous level in the menu and set the displayed values. The latest
settings are stored automatically and will be recalled when the menu is selected again.
EXIT will reset the firmware to the main screen.
All functions are indicated on the display; max. 8 lines and 40 characters.
EM TEST
UCS 500 M4
Ultra Compact Simulator
V 3.00a02
Burst
Surge
Power Fail
ESD
SWN: 001234
The serial number and the version number SWN are
used for traceability reasons. These numbers are listed
in the test reports and calibration certificates. These
numbers also are listed within the test reports
generated by the ISM ISO software.
Start-up display example UCS 500M4. The models type is displayed after startup.
4.2.
Menu structure
Level 0...4
Level 0
Level 1
Level 2
Level 3
Level 4
MAIN MENU
F1
BURST IEC 61000-4-4
F2
SURGE IEC 61000-4-5/9
F3
PFS
IEC 61000-4-11/8/29
F4
ESD
IEC 61000-4-2
F7
SERVICE
Burst
IEC 61000-4-4
F1
Quick Start
F2
Standard routines
F3
User test routines
Quick Start
F1
Start
F2
Change
F3
Continue
Start
Start the test routine
Change
Select all parameters
Continue
Continue the test routine
Surge
IEC 61000-4-5/9
Standard routines
Preprogrammed test routines as
per standard requirements
Standard routines
F1 : F4
IEC 61000-4-4 Level 1-4
F5
Generic Standard EN 61000-6-1
F6
Generic standard EN 61000-6-2
F7
Manual standard routine
Standard routines F1..F3
F1
Start
F2
Change
F3
Continue
Power Fail
IEC 61000-4-8/11/29
User test routines
Preprogrammed test routines for
evaluation and design support
F1
Synchronous to the mains
F2
Random burst release
F3
Change V
after T by
∆
U
F4
Frequency sweep I
F5
Frequency sweep
II
F6
Frequency sweep III
F7
Change polarity after T
User test routines F1..F3
F1 Start
F2 Change
F3 Continue
ESD
IEC 61000-4-2
Service
F1
Adresses
F3
Setup
F4
Change standard levels
Setup
F1
Change language
F2
LCD backlighting
F3
Interfaces
F4
ESD/keyboard beeper
F5
Running time clock
F6
Set voltage
F7
Magnetic field factors
Change language
Gern or English
LCD backlighting
On, Off or Auto
Interfaces
Select all parameters
ESD/keyboard beeper
(on, off)
Running time clock
Display of the TEST ON time
Set voltage
(ext. motor variac)
Magnetic field factors
Correction factors for magnetic field
antenna and current transformer
Change standard level
Change standard level
F1
All parameters to standard level
F2
IEC 61000-4
F3
EN 61000-6-1
F4
EN 61000-6-2
Change standard levels
F1
IEC 61000-4-4 Burst
F2
IEC 61000-4-5 Surge
...