11.1.2.
Standard Test Routines
Standard test routines
F1 : IEC 61000-4-11 (ac power supply mains)
F2 : IEC 61000-4-29 (dc power supply systems)
F3 : EN 61000-6-1
Generic standard
F4 : EN 61000-6-2
Generic standard
F7 : Manual test routine
F1
F2
F3
F4
F5
F6
F7
11.1.2.1.
F1: IEC 61000-4-11 ( AC power supply mains )
Page 3
Power Fail
IEC 61000-4-11
F1:
Level 1
0%
10.0 ms
F2:
Level 2
0%
20.0 ms
F3:
Level 3
40%
200 ms
F4:
Level 4
70%
500 ms
F5:
Level 5
80%
5000 ms
F6:
Short Interruptions
F7:
Voltage Variation
As long as the external variac MV2616
is used, controlled by an analogue 0-
10V control voltage, the test is
conducted automatically. If this option is
not available the manual test routine
shall be used.
F1
F2
F3
F4
F5
F6
F7
F1 ... F5: DIPS Level 1 to Level 5
Page 4 ( Show parameter)
Power Fail
IEC 61000-4-11
Level 3
V2% =
40%
A
= 0 dgr
td
=
200ms
tr
= 10s
n
= 3.
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7
F6: Short interruptions
Page 4 ( Show parameter)
Power Fail IEC 61000-4-11
Interruption
V2% =
0 %
A
= 0 dgr
td
=
5000ms
tr
= 10s
n
= 3
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7
F7: Voltage Variation
Page 4 ( Show parameter)
Power Fail IEC 61000-4-11
Variation
V2% =
70%
A
= 0 dgr
ts
=
20ms
tr
= 10s
ti
= 0.5s
n
= 3
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7
Attention:
This is a new Voltage Variation function which is required in the new FDIS IEC 61000-4-11. The Voltage Variation function
which is conform with the actual published IEC 61000-4-11 is available in the USER TEST ROUTINES.