11.6.
50/60Hz Magnetic Field as per IEC 61000-4-8
Magnetic field standard routine IEC 61000-4-8
<- -> O O
5 A/m 01:00 min
START O H/T
F1
F2
F3
F4
F5
F6
F7
With the inc knob the values for the magnetic field strength and test duration can be changed. The actual
parameter is indicated by a blinking circle.
Test setup
Warning
For magnetic field testing the power mains input at PF1 shall be
disconnected.; 230V/16A.
The voltage V is adjusted with a variac as long as the required antenna current is available and the related H field
is generated in the center of the magnetic field antenna.
The variac, type MV 2616 may be controlled automatically via the test generator. Any equivalent variac available
in the lab can be used to control the current manually. Please take care that the variac has a sufficient current
capability.
For more detailed information the operator can require an additional manual especially for magnetic field