9.1.3.
User Test Routines
The user can program, save and recall his own specific test routines. The next pages shows the selection of the
functions.
USER TEST ROUTINES
F1 : Synchronized
F2 : Random burst release
F3 : Voltage change after T by
F4 : Frequency sweep in one single burst
F5 : Frequency sweep with constant pulse numbers
F6 : Frequency sweep with constant burst duration
F7 : Change po/- after T
F1
F2
F3
F4
F5
F6
F7
After selection the last used test parameters will be indicated on the display.
Customized test routines
The software controls user test routines according to the specification of the user. All limitations are the same as
defined under Quick Start.
Synchronized with a fixed phase angle
The burst is triggered with respect to the phase angle of the
power supply connected to the Sync input at the rear panel
of the equipment. The power supply must be an AC voltage
with a nominal frequency of 16 to 500Hz. The phase must
be connected to L. This can be checked by the lamp
connected to the L input.
Random burst release
No repetition rate is selected. The single burst will be
triggered by statistics in the limits of 20 to 2000ms as time
between two bursts. All limitations are the same as defined
under Quick Start.
Voltage change after T by
∆
V The test voltage is increased
from V1 to V2 by steps of
∆
V after the defined test time T. All
limitations are the same as defined under Quick Start. The
limitation of the max. generated number of spikes is related
to the higher voltage of V1 or V2.