B-8 Pressure sensor
The pressure sensor can be implemented either as an inter nal or external variant.
Connection:
Pressure connection:
Ermeto M12 x 1.5 male thread
Usable length approx. 10 mm
Installation information:
When connecting the pressure line to the pressure sensor, attention must be paid to the
pipe diameter to avoid damage and leaks on the gland connection. In particular the parting
point on the pipe should be checked for burrs or flaring, which cause an i ncrease in the
outside diameter of the pipe (see following drawing of the Ermeto connection on the
pressure sensor).
Cross-section of pressure sensor connection
Measurement range
Max. admissible
overload
0,7 ... 2
bar abs.
18 bar abs.
0,8 ... 5 bar abs.
25 bar abs.
1,4 ... 7 bar abs.
25 bar abs.
2
... 10 bar abs.
40 bar abs.
2,4 ... 12 bar abs.
40 bar abs.
4 ... 20 bar abs.
40 bar abs.
6 ... 30 bar abs.
60 bar abs.
8 ... 40 bar abs.
60 bar abs.
14 ... 70 bar abs.
105 bar abs.
16 ... 80 bar abs.
105 bar abs.
B-9 Temperature sensor
Type:
Pt500 according to DIN EN 60751
Measurement range:
-10°C ... +60°C
Measurement uncertainty:
≤
± 0.1% of measurement
Mounting:
Insertion into sensor receptacle with variable length
B-10 Measurement uncertainty
The error limits quoted in the MID and in (DIN) EN 12405 are maintained.
Depending on the ambient temperature and the pressure measurement range, detailed
data is available on request.
Summary of Contents for 83 462 140
Page 117: ...Volume Conversion Device EK260 Elster GmbH 117 A 2 Approval for Ex Zone 1 ...
Page 118: ...Volume Conversion Device EK260 118 Elster GmbH ...
Page 119: ...Volume Conversion Device EK260 Elster GmbH 119 ...
Page 120: ...Volume Conversion Device EK260 120 Elster GmbH ...
Page 121: ...Volume Conversion Device EK260 Elster GmbH 121 ...