6. The diode at
ALARM
is activated, the display shows
0300
. Set the time delay, in seconds,
use
+
and
-
.
Press
=
to confirm.
9. Decide if the function shutdown at alarm shall be on (
Shon
) or off (
ShoF
) with
+
. Press
=
to confirm.
10. Step to
EXIT TO MEASURE
by pressing
M
repeatedly. Press
=
to start measuring.
Pulse width modulated regulation
4. The diode at
PWM
is activated, the display shows
100.0
. Set the total period time (in
seconds). Use
+
and
-
. Press
=
to confirm.
7. The diode at
P-BAND
is activated, the display shows
0100
.
Set the proportional band in %
of total range. Use
+
and
-
. Press
=
to confirm.
8. The diode at
I-TIME
is activated, the display shows
0050
.
Set the integration time (if no
integration is desired, set
0000
).
Use
+
and
-
. Press
=
to confirm.
9. The diode at
ALARM
is activated, the display shows
0300
. Set the time delay, in seconds,
use
+
and
-
.
Press
=
to confirm.
11. Decide if the function shutdown at alarm shall be on (
Shon
) or off (
ShoF
) with
+
. Press
=
to confirm.
11. Step to
EXIT TO MEASURE
by pressing
M
repeatedly. Press
=
to start measuring.
5.2.3 Output signal
1. Interrupt the measuring by pressing
+
and
-
at the same time until the measuring is
interrupted and the diode at
TEMP
is activated.
2. Step
to
OUTPUT
with
M
.
3. Choose between 0 - 20 and 4 - 20 mA, use
+
.
4. Press
=
to confirm.
5. The display shows
SPAn
. Press
=
to confirm.
6. The display shows
0000
. Set the %-value corresponding to 0 (alt. 4) mA.
7. Press
=
to confirm. The display shows
0100
.
8. Set the %-value corresponding to 20 mA. Press
=
to confirm.
9. The instrument steps automatically to
EXIT TO MEASURE
. Press
=
to start measuring.
NOTE! The values have to be set so that the 0/4 mA-level is corresponding to a lower value than the
20 mA value. In other cases the output signal stays at 0 mA.