Ultrasonic cleaning process
22
BA/ Transsonic TI-H_GB_0306
© Elma GmbH & Co KG
7.5
After the cleaning
As a rule the cleaned items must be rinsed and dried after the
cleaning process.
The choice of the rinsing medium or media depends on the
used cleaning media and on the cleanness requirements for the
cleaned items. It might be useful to rinse the items with
ultrasound.
Drain the liquid as soon as it is dirty or if the unit has not been
operated over a prolonged period of time (certain residues and
contaminations can damage the stainless-steel tank).
Use the quick-drain duct to empty the tank.. The bottom of the
transducer tank is inclined towards the drain to accelerate the
draining process.
For instructions on the cleaning of the transducer tank after
draining please see
section 9.1, Maintenance and care
.
Follow-up treatment
of cleaning items
Empty the unit
Cleaning of the
transducer tank