Before initial operation
14
BA/ Transsonic TI-H_GB_0306
© Elma GmbH & Co KG
5
Before initial operation
5.1
Unpacking and placement
If possible keep the original packing material for service
purposes.
Disposal must be in compliance with the valid regulations. You
can also return the packing material to the manufacturer or
supplier.
Check the unit for possible transport damages before initial
operation. In cas of visible damage do not operate the unit.
Contact your supplier and forwarding agent.
Place the unit on a solid, even and dry surface which is
resistant to possibly leaking cleaning liquid. Ensure that the
workplace is sufficiently ventilated!
Packing material
Check for transport
damages
Placement
Risk of electrocution due to humidity inside the unit!
Protect the unit from entering humidity.
The unit inside is splashwater-proof.
Keep workplace and housing dry in order to prevent electrical
accidents and damages on the unit.
CAUTION
For safe operation of the unit the following preconditions must
be guaranteed:
Ambient conditions
•
Allowed ambient temperature during operation:
+5°C to +40°C
•
Allowed relative humidity of air during operation: max. 80%
5.2
Connect the unit to the mains
TI-H 5 –
TI-H
25
: 1 phase (100 – 120 V or 220-240 V);
1 N; 1 PE protective earth.
TI-H 25 –
TI-H
160
: (200 – 208 V or 380 – 400 V)
3 phases 1 N; 1 PE protective earth.
Connect the unit to a grounded shockproof socket or a control
intended for this purpose.
Ensure that the values indicated on the nameplate of the unit
must correspond with the available connecting conditions.
Required mains
conditions
Connect mains cable