22
Figure 8.2 - Power Output vs Resistive Load
SURGITRON 4.0 DUAL RF/120 IEC
CUT MODE
NOTE:
The above power chart is a typical output characteristic of the Wolf /
ellman
Surgitron 4.0 Dual RF/120 IEC unit. Actual output power measurements may vary
with type of test equipment used and method of circuit connection. The actual
power as a function of load resistance and output control setting shall not deviate
by more than +/-20%