background image

5

cleaning or disinfecting the unit with flammable or explosive agents is unavoidable,
these must be completely evaporated from the HF Radio Surgical Generator before it
is switched on.

2.6.4.2

Handpiece and Electrodes

Cleaning:
Tissue residue can be removed from the active electrode with a sterile, moist cloth,
steel or copper wool. No scalpel, scissors or similar pointed objects must be used for
cleaning the electrodes, since they will damage the electrode surface. Such damage
increases adhesion of tissue to the surfaces of the electrodes during use.

Disinfecting:
Always clean and disinfect the electrodes in a disinfecting solution prior to sterilizing
them. All commercially available disinfectants may be used.

Sterilization:
Steam sterilize for 3 to 10 minutes for prevacuum, or gravity displacement (flash), at
132˚ to 135˚C (270˚ to 275˚F), 2.2 bar (32 PSI).

Disclaimer:

All autoclavable products have an indefinite shelf life. These items should be routinely inspected for any wear,

breaks, or deterioration of the insulation, and should be discarded and replaced if any of these common occurrences exist.

2.6.4.3

The dual footswitch is a waterproof unit, it can be cleaned with regular detergent.

2.6.5

The equipment is not suitable for use  in the presence of a FLAMMABLE ANESTHETIC MIXTURE
WITH AIR OR WITH OXYGEN OR NITROUS OXIDE.

2.6.6

The mode of operation is continuous operation with intermittent loading, duty cycle 10 seconds
on/30 seconds off.

2.6.7

The permissible operational environment conditions:
ambient temperature range

------ +10° C~ +40°C

relative humidity range

------ 30% ~ 75%

For transport and storage:
ambient temperature range

------ -10° C ~ +50°C

relative humidity range

------ 10%~95%

atmospheric pressure range

------ 500hPa~1060 hPA

This unit has been checked for proper and safe operation prior to shipment. However, all
equipment should be rechecked for proper and safe operation prior to initial operation, and before
each subsequent use.

2.7

PREVENTIVE INSPECTION AND MAINTENANCE

Regularly preventive inspection should be carried out to prevent reduced safety of the unit due to aging,
wear and tear, etc. The cable and electrode must be checked before each use.

The Wolf / 

ellman

Surgitron 4.0 Dual RF/120 IEC High-Frequency Radiosurgical Generator must be

inspected for technical safety at least once a year. These technical safety inspections mean:

• Inspection of electrical safety in compliance with IEC 60601-1 and 60601-2-2

• Inspection for damage to the unit and its accessories

• Inspection of proper function of the unit

ATTENTION: Alcohol must never be used to clean and disinfect the front panel. We

recommend that the panel be disinfected by spraying or wiping.
However, it is essential to follow the manufacturer’s directions
regarding cleaning, disinfecting and sterilizing of the accessories.

Summary of Contents for Surgitron 4.0 Dual RF/120 IEC

Page 1: ...074 01 03...

Page 2: ...1 Rear Panel 4 2 Front Panel V PREPARATION FOR USE Page 14 5 0 Preparation for Use 5 1 Operational Preparation and Functional Selection of the Surgitron Dual Frequency VI CLINICAL INFORMATION AND PRA...

Page 3: ...zor sharp knife Electrocoagulation is a nonvolatilizing destruction of tissue cells by a radiofrequency wave The atraumatic nature of electrosection provides a noteworthy advantage The lack of trauma...

Page 4: ...change without notice 2 1 GENERAL DESCRIPTION The Wolf ellman Surgitron 4 0 Dual RF 120 IEC enhanced capability Radiosurgical Generator described herein is a compact source of high frequency low temp...

Page 5: ...Deeply modulated square wave with average to peak ratio approximately 35 Modulation occurring at a 60 or 50 Hz rate 2 4 4 FULGURATE Deeply modulated wave with average to peak ratio to 10 or less Modul...

Page 6: ...WITH AIR OR WITH OXYGEN OR NITROUS OXIDE 2 6 6 The mode of operation is continuous operation with intermittent loading duty cycle 10 seconds on 30 seconds off 2 6 7 The permissible operational enviro...

Page 7: ...aintenance contract with the manufacturer or one of its authorized service agents is recommended after the warranty period 3 1 SAFETY INSTRUCTIONS Please review the below cautionary instructions to re...

Page 8: ...E The Wolf ellman Surgitron 4 0 Dual RF 120 IEC comprises some unique features which include retractable mode and power output level setting the waterproof dual footswitch for CUT and COAG modes and a...

Page 9: ...er down push button for CUT COAG mode l During this power adjustment the handpiece acts as two button fingerswitch handpiece thus the power output intensity may be tested m Depress the SELECT touch sw...

Page 10: ...should be set to a volume that can easily be heard by the user 4 1 5 Footswitch Receptacle This 9 pin receptacle accepts a Hypersonics Footswitch connector 4 1 6 Dual Footswitch Pedal The unit is ship...

Page 11: ...pment For error codes information see 7 5 Error Codes Table 4 2 2 Fault LED Indicator It will be Red if any error is detected otherwise it should be Green 4 2 3 Neutral Plate Receptacle The neutral pl...

Page 12: ...been selected See section 4 0 2 for Three Button Fingerswitch Handpiece application 4 2 12 Yellow Light Cut Coag Indicator The indicator illuminates when Cut Coag function has been selected See sectio...

Page 13: ...Indicator for Coagulation Mode The default state is zero when On Off power switch is turned on See section 4 0 1 for special features of the power storage The output power will remain at the selected...

Page 14: ...13 Figure 4 2 Numbers for Description of the Front Panel Control Elements...

Page 15: ...to select the desired power level Depress the CUT Footswitch controller or Three Button Fingerswitch to activate the output power See section 4 0 2 for three button fingerswitch handpiece application...

Page 16: ...ppreciate the tremendous advantages inherent in radiosurgery 6 1 1 Defining Good Technique Radiosurgery can create tissue damage if the heat is allowed to accumulate in the tissue to the point where e...

Page 17: ...de is then brought into contact with the hemostat 1 or 2 inches from its tips When the Partially Rectified Modulated Waveform is turned on the walls of the vessel will be sealed making ligation unnece...

Page 18: ...nt gentle contact with the tissue is performed until bleeding stops The same meat specimen used for cutting may be used for coagulating practice Effective coagulation is achieved when the treated area...

Page 19: ...IONS 7 0 TECHNICAL SPECIFICATIONS 7 1 OUTPUT CHARACTERISTICS 7 2 SAFETY FEATURES All waveforms are interrupted at a 37 Hz rate to allow sampling with the power off 7 2 1 Standard Compliance IEC60601 1...

Page 20: ...50 Hz 7 3 3 Power consumption at max output power 330 VA 7 3 4 Input current at max output power 1 30A 7 3 5 Rating of fuses in the main supply power T 2 0A 250V Two 7 4 SIZE AND WEIGHT 7 4 1 W x H x...

Page 21: ...IGITAL SETTING 500 RESISTIVE LOAD 8 2 CUT MODE POWER OUTPUT VS RESISTIVE LOAD 8 3 CUT COAG MODE POWER OUTPUT VS RESISTIVE LOAD 8 4 HEMO MODE POWER OUTPUT VS RESISTIVE LOAD 8 5 FULGURATE MODE POWER OUT...

Page 22: ...LAR MODES NOTE The above power chart is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used...

Page 23: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Page 24: ...art is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection...

Page 25: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Page 26: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load r...

Page 27: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load re...

Page 28: ...a com AUSTRIA Phone 01 405 62 19 Fax 01 405 51 51 45 Email info richard wolf at BELGIUM Phone 09 280 81 00 Fax 09 282 92 16 Email endoscopy richard wolf be FRANCE Phone 03 26 87 02 89 Fax 03 26 87 60...

Reviews: