8
a. “HOLD” should be displayed on the power display of CUT mode.
b. The power display of COAG mode starts to count down from “3”, “2”, “1”, then “0”.
Release the middle button.
c. Then, both power displays return to the previous power setting.
d. The blue light indicator is selected and "ON" for FOOTSWITCH MODE, which sets COAG
mode in standby waiting for the action from the fingerswitch controller.
e.
Yellow light (CUT) and (CUT/COAG) indicators will alternate "ON" and "OFF", they are in
standby, ready for action from the fingerswitch.
f.
Release the middle yellow button on the Three-Button Fingerswitch handpiece. The functions
are successfully activated.
To adjust the power output for CUTTING MODE, the procedure is:
g.
Depress the SELECT touch switch on the top yellow panel.
h. The "CUT" yellow light indicator stays "ON" instead of alternating with "CUT/COAG."
i.
Adjust the power output by depressing the power-up or power-down push button to the
desired power level for "CUT" mode.
j.
Depress the SELECT touch switch on the top yellow panel again to adjust CUT/COAG power
output, its yellow light indicator should be "ON" now.
k.
Adjust the power output by depressing power-up or power-down push button for "CUT/COAG"
mode.
l.
During this power adjustment, the handpiece acts as two-button fingerswitch handpiece, thus
the power output intensity may be tested.
m. Depress the SELECT touch switch again to reinstate the three-button fingerswitch function,
yellow lights of (CUT) and (CUT/COAG) should alternate "ON" and "OFF" now, and the blue
light of (HEMO) and (BIPOLAR) should alternate “on” and “off”.
Regression:
The three-button fingerswitch handpiece may be used as a conventional two-button fingerswitch
handpiece if the CUT/COAG mode is not activated. The Surgitron 4.0 Dual RF/120 IEC may be
operated as unit to adopt an IEC two-button fingerswitch handpiece, if a three-button fingerswitch
handpiece is not available.
4.1
THE REAR PANEL CONTROL ELEMENTS
see Figure 4.1 Rear Panel
4.1.1
Power Plug:
The Surgitron Unit is shipped with an approved hospital grade power connector. To ensure patient
safety, the Surgitron unit must be properly grounded.
4.1.2
Power Socket:
The Surgitron unit must be connected to correctly installed sockets with a grounded earth
conductor using the power cord supplied by the manufacturer.
4.1.3
Line Fuses:
The unit is protected with two line fuses. If a fuse blows, the unit should be checked for possible
faults before the fuse is replaced and the unit is operated again. Always replace with a fuse of the
same rating.