The active part of the electrode is always sterile when the Surgitron is in operative use. The tissue
will also be sterilized when it comes in contact with the active electrode. This is an extreme
advantage over the scalpel, which is contaminated by the nonsterile tissue surrounding the
surgical incision. The shank of the electrode, however, is not sterilized when the Surgitron is
activated. It is, therefore, suggested to either autoclave the electrodes or cold sterilize them. The
electrodes will have longer shelf life if they are cold sterilized. Do not attempt to operate with
electrodes if the protective insulation is cracked or worn. In the event that the brass is exposed, a
shock or burn may be felt by the operator or patient.

6.4

ANESTHESIA

Anesthesia, either local or general, must be used with radiosurgical procedures. IF NITROUS OXIDE
analgesia is used, a local anesthetic should also be used.

VII. TECHNICAL SPECIFICATIONS

7.0

TECHNICAL SPECIFICATIONS

7.1

OUTPUT CHARACTERISTICS

7.2

SAFETY FEATURES

All waveforms are interrupted at a 37 Hz. rate to allow sampling with the power off.

7.2.1

Standard Compliance

IEC60601-1

IEC60601-2-2

7.2.2

Neutral Plate Electrode

earthed at HF

7.2.3

Neutral Electrode monitor

MAX 1000 ohms 

resistance between the 2 contact 
areas of a split neutral electrode

Mode/

Output

Maximum

Power Output

Power Output

Activation

Character

Waveform

Output Power

vs. Setting

vs. Load Resistance

Cut

4.0 MHz CW

120W

See Figure 8.1

See Figure 8.2 

via foot pedal or

sinusoid

@ 500 

fingerswitch

Cut/Coag

4.0 MHz with

80W

See Figure 8.1

See Figure 8.3

via foot pedal or

rectified full-wave

@ 500 

fingerswitch

envelope

Hemo

4.0 MHz with 

60W

See Figure 8.1

See Figure 8.4

via foot pedal or

square wave 

@ 500 

fingerswitch

rectified e n v e l o p e

Fulgurate

4.0 MHz with

40W

See Figure 8.1

See Figure 8.5

via foot pedal or

modulated

@ 500 

fingerswitch

Bipolar

1.7 MHz with 1/2

120W

See Figure 8.1

See Figure 8.6

via foot pedal

wave rectified  

@ 200 

envelope

NOTE: The open-circuit Output peak-to-peak Voltage for all modes is approximately 1,000

Volts.

Summary of Contents for Surgitron 4.0 Dual RF/120 IEC

Page 1: ...074 01 03...

Page 2: ...1 Rear Panel 4 2 Front Panel V PREPARATION FOR USE Page 14 5 0 Preparation for Use 5 1 Operational Preparation and Functional Selection of the Surgitron Dual Frequency VI CLINICAL INFORMATION AND PRA...

Page 3: ...zor sharp knife Electrocoagulation is a nonvolatilizing destruction of tissue cells by a radiofrequency wave The atraumatic nature of electrosection provides a noteworthy advantage The lack of trauma...

Page 4: ...change without notice 2 1 GENERAL DESCRIPTION The Wolf ellman Surgitron 4 0 Dual RF 120 IEC enhanced capability Radiosurgical Generator described herein is a compact source of high frequency low temp...

Page 5: ...Deeply modulated square wave with average to peak ratio approximately 35 Modulation occurring at a 60 or 50 Hz rate 2 4 4 FULGURATE Deeply modulated wave with average to peak ratio to 10 or less Modul...

Page 6: ...WITH AIR OR WITH OXYGEN OR NITROUS OXIDE 2 6 6 The mode of operation is continuous operation with intermittent loading duty cycle 10 seconds on 30 seconds off 2 6 7 The permissible operational enviro...

Page 7: ...aintenance contract with the manufacturer or one of its authorized service agents is recommended after the warranty period 3 1 SAFETY INSTRUCTIONS Please review the below cautionary instructions to re...

Page 8: ...E The Wolf ellman Surgitron 4 0 Dual RF 120 IEC comprises some unique features which include retractable mode and power output level setting the waterproof dual footswitch for CUT and COAG modes and a...

Page 9: ...er down push button for CUT COAG mode l During this power adjustment the handpiece acts as two button fingerswitch handpiece thus the power output intensity may be tested m Depress the SELECT touch sw...

Page 10: ...should be set to a volume that can easily be heard by the user 4 1 5 Footswitch Receptacle This 9 pin receptacle accepts a Hypersonics Footswitch connector 4 1 6 Dual Footswitch Pedal The unit is ship...

Page 11: ...pment For error codes information see 7 5 Error Codes Table 4 2 2 Fault LED Indicator It will be Red if any error is detected otherwise it should be Green 4 2 3 Neutral Plate Receptacle The neutral pl...

Page 12: ...been selected See section 4 0 2 for Three Button Fingerswitch Handpiece application 4 2 12 Yellow Light Cut Coag Indicator The indicator illuminates when Cut Coag function has been selected See sectio...

Page 13: ...Indicator for Coagulation Mode The default state is zero when On Off power switch is turned on See section 4 0 1 for special features of the power storage The output power will remain at the selected...

Page 14: ...13 Figure 4 2 Numbers for Description of the Front Panel Control Elements...

Page 15: ...to select the desired power level Depress the CUT Footswitch controller or Three Button Fingerswitch to activate the output power See section 4 0 2 for three button fingerswitch handpiece application...

Page 16: ...ppreciate the tremendous advantages inherent in radiosurgery 6 1 1 Defining Good Technique Radiosurgery can create tissue damage if the heat is allowed to accumulate in the tissue to the point where e...

Page 17: ...de is then brought into contact with the hemostat 1 or 2 inches from its tips When the Partially Rectified Modulated Waveform is turned on the walls of the vessel will be sealed making ligation unnece...

Page 18: ...nt gentle contact with the tissue is performed until bleeding stops The same meat specimen used for cutting may be used for coagulating practice Effective coagulation is achieved when the treated area...

Page 19: ...IONS 7 0 TECHNICAL SPECIFICATIONS 7 1 OUTPUT CHARACTERISTICS 7 2 SAFETY FEATURES All waveforms are interrupted at a 37 Hz rate to allow sampling with the power off 7 2 1 Standard Compliance IEC60601 1...

Page 20: ...50 Hz 7 3 3 Power consumption at max output power 330 VA 7 3 4 Input current at max output power 1 30A 7 3 5 Rating of fuses in the main supply power T 2 0A 250V Two 7 4 SIZE AND WEIGHT 7 4 1 W x H x...

Page 21: ...IGITAL SETTING 500 RESISTIVE LOAD 8 2 CUT MODE POWER OUTPUT VS RESISTIVE LOAD 8 3 CUT COAG MODE POWER OUTPUT VS RESISTIVE LOAD 8 4 HEMO MODE POWER OUTPUT VS RESISTIVE LOAD 8 5 FULGURATE MODE POWER OUT...

Page 22: ...LAR MODES NOTE The above power chart is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used...

Page 23: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Page 24: ...art is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection...

Page 25: ...t is a typical output characteristic of the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection T...

Page 26: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load r...

Page 27: ...the Wolf ellman Surgitron 4 0 Dual RF 120 IEC unit Actual output power measurements may vary with type of test equipment used and method of circuit connection The actual power as a function of load re...

Page 28: ...a com AUSTRIA Phone 01 405 62 19 Fax 01 405 51 51 45 Email info richard wolf at BELGIUM Phone 09 280 81 00 Fax 09 282 92 16 Email endoscopy richard wolf be FRANCE Phone 03 26 87 02 89 Fax 03 26 87 60...

Reviews: