88
Section 1: Introduction
WARNING!
If power is lost while the stainer is running, the lid will remain locked until power is restored. Do not
attempt to open the lid while power is off.
WARNING!
Electrical shock hazard: Do not open this instrument or attempt internal repairs. Refer servicing to
qualified service personnel. Contact ELITechGroup Biomedical Systems service.
CAUTION:
This instrument has been designed and tested to CISPR 11 Class A and FCC Part 15 Class A. In
a domestic environment it may cause radio interference, in which case, measures to mitigate
the interference may be necessary. This instrument complies with the emission and immunity
requirements described in the IEC 61326 series.
In an electromagnetic environment, the environment should be evaluated prior to operation of the
device.
Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g.
unshielded intentional RF sources), as these may interfere with the proper operation.
CAUTION:
To avoid serious instrument damage, always use reagents supplied by ELITechGroup. Using
reagents not supplied by ELITechGroup may void the warranty.
CAUTION:
Only spare parts supplied or specified by ELITechGroup should be used in this instrument. Using
non-approved parts may affect the performance and safety features of the instrument. If the
instrument is used in a manner not specified by ELITechGroup, the protection provided by the
instrument may be impaired. If in doubt, contact your ELITechGroup representative.
Functional Description
The Aerospray Cytology Slide Stainer/Cytocentrifuge (Model 7522) is a dual-purpose, microprocessor-
controlled slide staining and cell preparation system. In use, atomizing spray nozzles apply fresh reagents
onto microscope slides prepared with cytological specimens. The slides are mounted in a rotating carousel
for processing.
Staining options include a Progressive and Regressive stain mode with programmable stain settings for
user customization.
Key Features
• Minimized reagent consumption
• Rapid staining
• Barcode scanner for tracking specimens and reagents
• Reagent and specimen traceability
• User traceability
• Administrator password
• Interactive touchscreen display