Section 7: Solving Problems
Reagent Level Detect System Errors
Reagent A-E Not Calibrated
During the second part of calibration, if no bottles are detected, the display shows an error message.
Calibrate again, making sure that the reagent bottles are inserted in those tray positions that have been
enabled in the level detection system.
LD (Level Detection) Unstable
If movement was detected on bottles while calibrating/zeroing, the display shows an error message.
NOTE:
While zeroing or calibrating, do not bump the instrument or lab bench. Ensure that no nearby
equipment vibrations can be transmitted to the stainer.
Calibrating the Reagent Level Detect System
If the Reagent Level Detect System is reporting incorrectly and zeroing (Section 2.2) does not correct the
problem, calibrate the system as follows:
1. Press
System Information
from the Main menu.
2. Press
System Setup
.
3. Press Level Detect.
4. Press
Calibrate
. Follow the display prompts.
5. Remove all reagent bottles and press
Start
.
NOTE:
Any vibrations or bumps to the instrument or lab bench can cause inaccuracies in zeroing or
calibration.