Product application, installation and service manual 

 

 

 

 

 ELIS PLZE

Ň

 a. s. 

Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 

Page 26 of 44 

 

ELIS PLZE

Ň

 a. s., Lu

č

ní 15, P. O. BOX 126, 304 26 Plze

ň

, Czech Republic, tel.: +420/377 517 711, fax: +420/377 517 722 

Es 90 337 K/c

 

6.2.2.8 Resetting aggregate quantities 

 

Upon  entering  this  mode,  the  corresponding  message  will  appear  on  the  display  (see  Fig.  25).  If  resetting  of 

aggregate quantities is not required, depress push-button T2 to access the next parameter-setting block. Confirm your 
intention to reset the aggregate quantities by depressing T4. The display will then ask  

 
    

           

 

 

 

RESET ?  

 

At this stage, you can still return to the initial step of the resetting mode by depressing T3. Depress T4 to reset the 

aggregate readings of the flow volume and flow mass, the operational information (the meter operation time, the error 
time and power loss time) and define the start of a new measurement period (the date, hour and minute of the same). 
The system will respond with a confirming message (Parameter Set). 

6.2.2.9. Meter zero setting 

 

Before  leaving  the  manufacturing  plant,  every  flow  meter  is  carefully  set  for  correct  operation.  One  of  the  key 

parameters in this respect is the meter zero. A correctly set meter zero implies that at zero flow rate (or zero fluid flow 
velocity through the meter sensor) the meter indicates a zero flow rate (zero fluid flow velocity). The setting value (a 
meter zero shift) is expressed in mm per second. The meter zero shift as identified in the manufacturing plant is stored 
in the meter memory under the name of initial (in-production) zero setting value. 

 

Meter component ageing and other factors acting over long-term operational periods may result in minor meter zero 

displacements. To eliminate these, use the automated zero resetting function. However, great care should be taken in 
employing this function. First of all, the zero flow rate condition shall be ensured (make sure that the closing valve in the 
piping is not leaking). Only then the zero resetting function may be used. 

 
A detailed description of the zero resetting block is shown in Fig. 26. Upon initiating this function, the operator shall 

select  either  the  in-production  or  automated  meter  zero  setting  mode.  The  selection  is  done  by  push-button  T1, 
confirmation by T4. When the in-production setting mode is selected, the meter zero is reset using the zero shift value 
determined in the manufacturing plant. 
 

In the automated zero setting mode, the meter will first ask whether the fluid flow rate through the meter sensor is 

really zero (the main requirement for a successful zero setting). If it is not so, cancel the setting process using push-
button T3. Upon confirmation by push-button T4 the display will show the message “WAIT FOR 100”. The zero setting 
procedure  lasts  100  measuring  cycles.  The  actual  number  of  measuring  cycles  performed  is  shown  on  the  second 
display line. 

 
After 100 measuring cycles the zero displacement is evaluated. If it is smaller than 50 mm/s, the shift value is stored 

and  the  display  with  read  “PARAMETER  SET”.  If  the  value  is  greater  than  50  mm/s, 
a notice to this effect is displayed. However, this is highly unlikely; in such as case it is recommended to check again 
whether the fluid flow  has indeed  been completely  stopped. Use  push-button T3 to invalidate the  setting  and  push-
button T4 to run the setting procedure anew. 

The meter zero setting function is available only with technological flow meters. 

6.2.2.10 End of parameter setting 

 

At the end of the parameter setting procedure, the display will read  

 
    

           

 

 

PAR. SETTING END 

 

 
Depress  push-button  T4  to  access  the  data  display  mode.  However,  should  you  wish  to  perform  any  additional 

parameter setting action, depress T3 to return to the previous parameter setting function block (see Fig. 27). 

 
 
 

Summary of Contents for SONOELIS SE 4015

Page 1: ...anual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 1 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c U...

Page 2: ...1 3 Display of measured data 13 6 1 2 Review of the measured quantities 13 6 1 3 Review of the measured quantity units 13 6 1 4 Unit conversion table 13 6 2 Keyboard control functions 13 6 2 1 Data d...

Page 3: ...ow Meter SONOELIS SE 4015 and SE 4025 Page 3 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c 8 TESTING 13 9 ORDERING 13 10 PACK...

Page 4: ...the ultrasonic probes the ultrasonic ray is sent in turns in and against the fluid flow direction 3 TECHNICAL DESCRIPTION 3 1 General information The SONOELIS SE 4015 SE 4025 ultrasonic flow meter is...

Page 5: ...the fluid flow in both directions and indicate the actual fluid flow direction On request the meter sensor can be supplied in an IP 68 housing 3 2 Design specifications 3 2 1 Sensor dimensions L O D D...

Page 6: ...ted to the piping the main pipe section and two pipe branches holding the ultrasonic probes see Fig 4 In the standard sensor version the body is designed for operating pressure PN 10 made of high qual...

Page 7: ...ets The grommets are intended for tight fitting of cables of external diameter 6 to 8mm PG 9 and 4 to 6mm PG 7 At the bottom of the box there is also an earthing bolt Both the front panel and the term...

Page 8: ...5 78 75 100 156 3 200 Q1 Minimum flow rate G min 44 68 8 88 1 110 138 173 220 275 347 440 688 881 m3 h 16 25 32 40 50 63 80 100 126 160 250 320 Q2 Transient flow rate G min 70 4 110 141 176 220 277 35...

Page 9: ...for flow rate Q Q2 see Table 2 and fluid temperature up to 50 C SE 4015 0 5 for flow rate Q Q2 and fluid temperature up to 50 C SE 4025 Fluid temperature 0 150 Ambient temperature 0 50 Ambient relati...

Page 10: ...nsor are 5DN and 3DN respectively see Fig 7 FLOW DIRECTION 3 DN 5 DN 5 DN 3 DN FLOW DIRECTION Fig 7 Minimum lengths of straight piping sections Fig 8 Minimum length of straight piping section at the i...

Page 11: ...ons where there is a piping outlet into open reservoirs or tanks anywhere near see Fig 12 FLOW DIRECTION FLOW DIRECTION Fig 12 Examples of incorrect sensor placement Another factor that may influence...

Page 12: ...Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c 3 0 3 0 PROBE 1 PROBE 2 Fig 13 Permitted range of one beam sensor rotation with respect to its long...

Page 13: ...atic and mechanical stresses by means of suitable installation boxes All cables shall be lead outside thermal insulation layers on piping if any Connections to the Pt 100 temperature sensor current ou...

Page 14: ...ecifications on the flange circumference Unless required otherwise the sensor shall be supplied with flanges according to standard SN EN 1092 1 the alternative solutions are ANSI or JIS flanges 5 2 3...

Page 15: ...puts and the RS 485 communication interface The X4 CANON 9 connector serves the purpose of connecting the RS 485 interface used in the meter calibration servicing and in company operational setting Th...

Page 16: ...tion on the display unit Combination of switch positions Measurement Instantaneous flow rate Measurement Instantaneous velocity Programming EEPROM programming Servicing Service The servicing mode is r...

Page 17: ...temperature range Rated pressure PN Fluid flow direction Protection class information 5 2 5 Sealing of technological meters The SONOELIS flow meters are provided with clamping and stick on factory and...

Page 18: ...ed to show the volume flow rate in m3 hod or the mass flow rate in metric tons per hour and the second display line to display the total volume in m3 or the total mass in metric tons alternatively wit...

Page 19: ...s l kg l hour kg hour 1 000 bbl 1 000 tons l min kg min bbl ton l s kg s 1 000 ft3 lb bbl hour tons hour ft3 bbl min tons min 1 000 gal bbl s tons s gal ft3 hour lb hour ft3 min lb min ft3 s lb s gal...

Page 20: ...articular operating conditions at the user s plant The push buttons T1 to T4 are provided with the following graphic symbols T1 T2 T3 T4 The push button control of the meter is shown in a schematic di...

Page 21: ...is determined by the manufacturer with respect to the meter application Provided the flow meter has been set for measurements in both directions of flow the sign before the reading indicates the flow...

Page 22: ...flow in the reverse direction 6 2 1 11 Temperature The temperature reading is only available if a thermometer is included in the meter configuration The reading sensitivity is 0 1 C 6 2 1 12 Density...

Page 23: ...ncorrect password the system will request a new password entry After three consecutive entries of incorrect passwords the system will switch over to the data display mode and will not permit further e...

Page 24: ...ed language Upon selecting the language confirm the setting by depressing push button T4 The message on the display will inform the operator of completion of the parameter setting in the newly selecte...

Page 25: ...be recorded as 03 07 01 and the time 7 minutes past 9 a m as 09 07 00 Confirm the selection by depressing push button T4 In the case of a commercial invoicing meter neither qs impulse number or sensi...

Page 26: ...ion However great care should be taken in employing this function First of all the zero flow rate condition shall be ensured make sure that the closing valve in the piping is not leaking Only then the...

Page 27: ...appear on the display LIQUID DOES NOT FLOW Check the zero flow rate condition and depress and release S1 The test will continue by checking whether the passage route for the ultrasonic ray in one dire...

Page 28: ...e meter needs be put out of service and either sent for repair to the manufacturing plant or a service technician be asked to come and repair the meter on site Should a fault be indicated at the ultra...

Page 29: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 29 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Page 30: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 30 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Page 31: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 31 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Page 32: ...lation and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 32 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 51...

Page 33: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 33 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Page 34: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 34 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Page 35: ...ncluding natural disaster A requirement for warranty repair shall be communicated to the manufacturer by a notice in writing by e mail fax or registered letter Should the manufacturer find the warrant...

Page 36: ...015 system optional assembly and testing fixtures this manual a statement on the product compliance and delivery note 12 WARRANTY CONDITIONS Unless agreed otherwise between the manufacturer and custom...

Page 37: ...ORDER NUMBER S E 4 0 5 Sensor design one beam sensor 1 double beam sensor 2 SPECIFICATION Dimension 200 800 0 1 of sensor 250 1250 0 2 DN nominal 300 1600 0 3 flow rate Q3 350 2000 0 4 400 2500 0 5 4...

Page 38: ...3 0 1 for remote 5 0 2 version m 7 0 3 10 0 4 15 0 5 20 0 6 30 0 7 40 0 8 50 0 9 60 special cabel 1 0 70 special cabel 1 1 80 special cabel 1 2 90 special cabel 1 3 100 special cabel 1 4 non standard...

Page 39: ...Q4 4 20 mA 3 non standard 0 Binary not required 0 1 output fluid flow direction upstream switch on 0 2 fluid flow direction upstream switch off 0 3 total fluid volume level on exceeding switch on 0 4...

Page 40: ...l min 0 6 l s 0 7 bbl h 0 8 bbl min 0 9 bbl s 1 0 ft3 h 1 1 ft3 min 1 2 ft3 s 1 3 gal h 1 4 gal min 1 5 gal s 1 6 Q4 1 7 non standard 0 0 Units of volume is not pictured 1 volume m 3 2 m 3 10 3 3 l 4...

Page 41: ...ver lich sud ls 3 O of flow sud sud ss 4 M meter lich lich ll 5 U Group 000 0 0 0 N 001 0 0 1 I 002 0 0 2 C A T I 255 2 5 5 O Address 000 0 0 0 N 001 0 0 1 002 0 0 2 255 2 5 5 CALIBRATION Calibration...

Page 42: ...47 48 49 50 51 52 ORDER NUMBER DELIVERY TERMS Number of 1 piece 0 0 1 pieces 2 pieces 0 0 2 3 pieces 0 0 3 999 pieces 9 9 9 1000 pieces or more 0 0 0 Packing no packing 1 standard 2 export 3 non stan...

Page 43: ...tallation and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 43 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377...

Page 44: ...SE 4025 Page 44 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c Manufacturer s address ELIS PLZE a s Lu n 15 P O BOX 126 304 26...

Reviews: