8.
TROUBLE SHOOTING
If a problem occurs, the unit will normally sound an alarm and the respective icon will flash. The
audible alarm can be silenced by pressing the mute button.
The appropriate action to then take (for the more common alarm displays) is shown on the Quick
Reference Guide (fitted to the inside of the door of the
MEDICA EDI - US.
)
If the unit cannot be
repaired using the information shown, please call your local ELGA LabWater representative
(See
Section 12.0 - Useful Contact Details).
WARNING!
ALWAYS ENSURE THAT THE MAINS POWER SUPPLY
IS ISOLATED BEFORE WORKING INSIDE THE UNIT.
Alarm Summary
Display
id
icon
Alarm
Recommended Action
672-98
Leak
Detection
PRESS mute to silence alarm
REMOVE power rectify leak and dry sensors
Dry the contacts
POWER on to reset.
672-97
Reservoir
level
controls
PRESS mute to silence alarm
REMOVE power
Contact Technical Support
672-96
High loop
pressure
PRESS mute to silence alarm
REMOVE power
Contact Technical Support
672-95
Permeate
over
temperature
PRESS mute to silence alarm
Potable feed temperature is unacceptably high >50
C
Contact Technical support
672-94
Recirc. loop
over
temperature
PRESS mute to silence alarm
Water temperature is unacceptably high >50
C
Contact Technical support
-93
LC177
Protek L2
incorrectly
installed
PRESS mute to silence alarm
CONFIRM Protek is correctly installed
POWER OFF/ ON to reset
-92
LC207
LC207
incorrectly
installed
PRESS mute to silence alarm
CONFIRM Medpure L1is correctly installed
POWER OFF/ ON to reset
-90
1M
Ω.cm
Water purity
alarm
PRESS mute to silence alarm
CONFIRM purity displayed is suitable for application.
CHANGE LC207 if water purity is insufficient
-89
35°C
Water
temperature
alarm
PRESS mute to silence alarm
CONFIRM temperature is suitable for application.
ADJUST alarm setpoint or DIRECT water to drain to
Introduce cool water
-88
100µS/cm
Permeate
purity alarm
PRESS mute to silence alarm
CONFIRM purity displayed is suitable for application.
ADJUST alarm setpoint or contact local service
provider.
-87
35°C
Permeate
temperature
alarm
PRESS mute to silence alarm
CONFIRM temperature is suitable for application.
ADJUST alarm setpoint or confirm potable supply
temperature is suitable
-86
Break tank
low
PRESS mute to silence alarm.
CONFIRM pressure and flow of potable supply
If problem persists contact local service provider
-85
UV lamp fail
PRESS mute to silence alarm
CHANGE lamp at the next opportunity
60
Invalid C.C.
Check if you have re-installed old LC206
Put in new LC206
57
Replace
LC206
LC206 is expired. Replace with new one.
56
LC206
reminder
Make sure you have a new LC206 ready in stock for
replacement 10% of life left.