MEDICA EDI
– US Operator Manual
ELGA
Page 28
MEDICA EDI
–US MANU40685 Ver 2 – 10/15
6.2
Tablet sanitization
The RO is sanitized to reduce the growth of microbiological
contamination within the RO module. The
MEDICA
has a built-
in reservoir which allows the sterilant to be introduced into the
water feeding the RO. Please read this entire section to
become familiar with the procedure before you start.
The sanitization frequency required is dependent on the
feedwater, local environment, usage patterns and application.
As a general rule, ELGA LabWater recommends sanitization of
the RO at 6 monthly intervals. However, the period between
sanitizations could be extended in particular circumstances.
For example, microbial growth will usually be lower with clean
feed water, well-maintained pre-treatment, low temperatures
and heavy usage.
WARNING! ALWAYS WEAR RUBBER GLOVES,
APRON
AND
FACEMASK.
DO
NOT
BREATHE FUMES OR ALLOW TO COME
IN CONTACT WITH SKIN OR EYES.
ALWAYS FOLLOW RECOMMENDATIONS
FOUND
IN
THE
MANUFACTURERS
MATERIAL SAFETY DATA SHEET AND
ANY APPLICABLE OSHA STANDARDS
FOR THE CHEMICAL BEING USED.
WARNING! LABEL
THE
MACHINE
WITH
APPROPRIATE WARNING SIGNS SUCH
AS
“DO
NOT
USE/
CONTAINS
STERILANT” (NOT PROVIDED)
6.2.1 RO sanitization (Step 1).
Sanitization of the finished installation is essential to ensure that
the system is properly commissioned and capable of achieving the
bacteriological control required for a typical application.
Sanitization selection screen
1.
PRESENT GREEN sanitization key to reader.
2.
PRESS
✓ to enter sanitization.