Page 35
10.
TECHNICAL SPECIFICATION
Feedwater
Source
Potable tap water as detailed below.
Failure to comply with the minimum feedwater pretreatment recommendations
will affect the life and performance of key components within the
MEDICA EDI -
US
, and may invalidate the warranty.
Contaminant
Measure
Range
Pretreatment
Calcium
Ca ppm as CaCO
3
< 250
None
> 250
Softener or
use very low RO recovery
Free chlorine
Cl ppm
< 0.1
Protek L2
0.1 to 0.5
Protek L2
<0.5 ppm LC123
> 0.5
Protek L2
Chloramine
Chloramine ppm
<0.05
LC123
0.5 to 1.0
Protek L2 if water usage is about
5000 litres per week.
>1.0
Cylinder of carbon sized correctly.
Silica
SiO
2
ppm
< 30
Protek L2
> 30
20 inch cartridge depth filter*
<30
LC123
Fouling Index
FI
<10
Protek L2
>10
<3
Backwashable media filter with a
minimum flow rate of 20l/min
LC123
Iron/manganese
Fe/Mn ppm
< 0.05 to 0.1
Protek L2, LC123
> 0.1
Back-washable pre- filter *
Organics
TOC ppm C
<3
Protek L2
> 3
Cylinder of carbon sized correctly
TEMPERATURE
4 - 40ºC (Recommended 15 - 25ºC)
FLOWRATE
(maximum requirement @15
O
C)
9l/min
Drain requirements (gravity fall with air gap).
20l/min
Feedwater Pressure
6 bar (90 psi) maximum, 2 bar (30 psi) minimum.
*
Installed in feed water supply.
Note: If feedwater purity is variable or values are
close to the top of one of the ranges, provide
pre-treatment for the higher range or seek
advice from Technical Support at ELGA
LabWater.
Pre-treatment LC177 and LA727 (LC123)
included in starter kits LC238 and LC237
respectively.