15. Annexe
15.1 Error messages and trouble shooting
Error message
E 1:
Sensor not compatibe.
E 2:
No sensor connected. This error message is displayed when no
sensor is connected after a total reset.
E 3:
Sensor defective. Displayed either when gauge is switched on or
if the sensor has been disconnected during operation.
E 4:
Sensor supplies instable measuring values (for example due to
strong magnetic fields in the surroundings or when measuring on
very soft material).
E 5:
Sensor too close to metal when gauge was switched on.
E 6:
Battery voltage low. .
E 11:
Data memory saturated. .
E 12:
Zero calibration not permitted. .
E 13:
CTC- calibration not permitted.
E 14:
2 point calibration (using two calibration foils) calibration not
permitted.
E 15:
1 point calibration not possible after CTC calibration.
E 16:
The application selected is already assigned to another sensor or
sensor had been repaired.
E 17:
Sensor not initialized
E 18:
BT-Device not found
E 19:
BT-Connection -Fail
E 20:
CALIBRATION FAIL
E 21:
SD Card FAULTY
E 22:
FILE ERROR
Following troubles can be remedied by carrying out a total reset (see Section 11.1).