14. Technical data
© ElektroPhysik
MiniTest 2500 - MiniTest 4500
63 of 73
N-Sensors (Eddy current principle)
Sensor
Range
Low range
resolution
Measuring
uncertainty on
ElektroPhysik
standards
(laboratory cond.
Min.Radius of
curvature
(convexe/
concave)
Min.
measuring
area
Min. substrate
thickness
Dimensions
in mm
N.08 Cr
0...80 µm
0,1 µm
± (1% of reading
+ 1 µm)
2,5 mm / 10 mm
Ø 5 mm
≥
100 µm for
Cu on Fe
Ø 16 x 99
N 02
0...200 µm
0,1 µm
± (1% of reading
+ 0,5 µm)
1 mm / 5 mm
Ø 0,2 mm
50 µm
Ø 16 x 99
N 1.6
0...1600 µm
0,1 µm
± (1% of reading
+ 1 µm)
1,5 mm / 10 mm
Ø 5 mm
50 µm
Ø 15 x 22
N1.6/90 internal
tube sensor
0...1600 µm
0,1 µm
± (1% of reading
+ 1 µm)
flat / 6 mm
Ø 5 mm
50 µm
8 x 11 x 180
N2/90
internal tube
sensor
0...2000 µm
0,2 µm
±( 1% of reading
+ 1 µm)
flat / 6 mm
Ø 5 mm
50 µm
8 x 11 x 180
N10
0...10mm
10 µm
± (1% of reading
+ 25 µm)
25 mm / 100 mm
Ø 50 mm
50 µm
Ø 60 x 50
N20
0...20 mm
10 µm
± (1% of reading
+ 50 µm)
25 mm / 100 mm
Ø 70 mm
50 µm
Ø 65 x 75
N100
0...100 mm
100 µm
± (3% of reading
+ 0,3 mm)
100 mm / flat
Ø 200 mm
50 µm
Ø 126 x 155
CN02
Cu-coatings on
insulating
substrate
10...200 µm
0,2 µm
± (3% of reading
+ 1 µm)
Only on flat
surfaces
Ø 7 mm
none
Ø 17 x 80