PS 2000 B Triple Series

3.4.8  Tracking mode

Tracking is sort of a master-slave operation of output 1 and 2 and where output 1 would be the master. This is why 

the tracking mode can only be activated and deactivated on the HMI of output 1 (left side).
As long as tracking is activated, the right-hand HMI of output 2 is completely locked and all set values and settings 

are submitted to output 2, so it follows output 1. Tracking is especially helpful when wanting to connect output 1 

and 2 in either series or parallel connection, in order to achieve higher output voltage or current.
Following applies to tracking:

• 

Tracking can only be activated when both outputs are off and none is in remote control

• 

Activated tracking is saved along with other device settings and thus instantly restored after powering the device

• 

Even in tracking mode, the two outputs can have separate alarms, which would always lead to a switch-off of 

both outputs; since alarms need to be acknowledged, the outputs can only be switched on again if both displays 

don’t show any alarm anymore

• 

Tracking can also be activated, deactivated and controlled remotely

Depiction of tracking mode in the displays:

Left-hand display:

Normal handling

Right-hand display:

HMI completely locked 

 

How to activate/deactivate the tracking mode

1. 

While the DC outputs 1 and 2 are switched off, press the 

Settings

 button of the left-hand HMI.

2. 

In the menu use the left knob to navigate to 

Tracking

 and press the knob.

3. 

In menu 

Tracking

 use the knob to activate tracking with 

On

 or deactivate with 

Off

. The selection of 

On

 might 

be blocked by any of the above listed reasons.

The mode, if activated anew, will become active after exiting the setup menu. Both display should then indicate 

Tracking

, as shown in the figures above.

Summary of Contents for 39 200 120

Page 1: ...Elektro Automatik PS 2000 B TFT Triple DC Power Supply Operating Manual Doc ID PS2TEN Revision 14 Date 11 2020...

Page 2: ......

Page 3: ...res before installation and use 14 2 3 2 Preparation 14 2 3 3 Installing the device 14 2 3 4 Connection to DC loads 15 2 3 5 Grounding of the DC outputs 15 2 3 6 Connecting the USB port 15 3 OPERATION...

Page 4: ...as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for general safety notices instructions and damage protection bans S...

Page 5: ...B Construction Version B Second generation Maximum current of the device in Ampere Maximum voltage of the device in Volt Special marking 3 Three outputs triple Series 2 Series 2000 Type identification...

Page 6: ...ulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device is switched off It s not necessary...

Page 7: ...he responsibilities for operation maintenance and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularl...

Page 8: ...pair of DC output sockets and is called output 2 The upper pair of DC output sockets form output 3 The meaning of the signals on the display of output 1 and 2 is as follows Signal OT OverTemperature...

Page 9: ...P 24 mVRMS Output 1 2 Current Adjustment range 0 6 A 0 10 A 0 3 A 0 5 A Accuracy 1 at 23 5 C 0 3 IMax 0 3 IMax 0 3 IMax 0 3 IMax Load regulation at 0 100 UOUT 0 15 IMax 0 15 IMax 0 15 IMax 0 15 IMax O...

Page 10: ...Automatik GmbH Helmholtzstr 31 37 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 2000 B Triple Series 1 8 4 Views Figure 1 Front side...

Page 11: ...alue or adjustment of parameter value in the menu Push selection of the decimal position cursor of the currently assigned value 7 Pushbutton On Off Used to switch the DC output oon or off unless the H...

Page 12: ...es For these devices the following accessories are available Multi Control license Ordering numbers Single license 33 100 229 5 pack of licenses 33 100 230 The device is delivered with an USB stick th...

Page 13: ...tatus display lower part This area displays various status texts and symbols Display Description The HMI is locked The HMI is unlocked Remote The device is under remote control via USB Alarm xxx Alarm...

Page 14: ...3 1 Safety procedures before installation and use Before connecting to the mains ensure that the connection is as shown on the product label Overvoltage on the AC supply can cause equipment damage In...

Page 15: ...ding of the DC outputs The metal socket in proximity to output 1 is internally wired to PE and can be used to ground one of the DC output poles of all three outputs by direct connection or it can be u...

Page 16: ...e Voltage regulation is also called constant voltage operation CV The DC output voltage of a power supply is held constant on the adjusted value unless the output current reaches the adjusted current...

Page 17: ...will be in constant voltage mode If however the actual current reaches the set value of current the device will switch automatically to current limiting While the DC output is switched on and constan...

Page 18: ...ltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage higher than set for the overvoltage alarm threshold OVP 0 110 UNo...

Page 19: ...or a blackout doesn t leave the device enough time to store the latest values It s practically powered down immediately Thus the device stores the set values and all settings automatically when they...

Page 20: ...left HMI Element Description In right HMI Language Selection of the display language between German and English Default setting English No Backlight The choice here is whether the backlight remains Al...

Page 21: ...bs 1 With the main display being shown it means no menu mode is active turn the left hand knob to adjust the output voltage and the right hand knob to adjust the output current This doesn t depend on...

Page 22: ...How to lock the HMI with PIN Don t activate the PIN lock if you are unsure about the current PIN It can be changed but only if the current PIN is entered Resetting the device via remote command will a...

Page 23: ...ivated tracking is saved along with other device settings and thus instantly restored after powering the device Even in tracking mode the two outputs can have separate alarms which would always lead t...

Page 24: ...t used the serial current is limited by the output with the lower current set value Series connection of multiple devices isn t permitted because the total voltage could exceed the voltage withstand o...

Page 25: ...considered completely differently to the other outputs It s meant to be an auxiliary output with a fixed voltage which is barely adjustable ca 3 to ca 6 V with a trimmer That trimmer is placed between...

Page 26: ...can try to acknowledge and delete the alarm by pressing button On Off In order to acknowledge an alarm during remote control see the programming guide These device alarms are configurable Alarm Meani...

Page 27: ...detail as possible is attached if shipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by a fusible which is inside a f...

Page 28: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 37 41747 Viersen Germany Fon 49 2162 37 85 0 Telefax 49 2162 16 230 ea1974 elektroautomatik de www elektroautomatik de...

Reviews: