Technical data
1
Guidance and manufacturer's declaration – electromagnetic immunity
The Elekta Neuromag
®
system is intended for use in electromagnetic environmen specified below. The
customer or the user of the Elekta Neuromag
®
system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %. An ESD
preventing floor covering is
recommended.
Electrical fast
transients /
burst
IEC 61000-4-5
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital
environment. The system is equipped
with no-voltage mains release switch;
after mains interruption a normal power-
up sequence by the operator is
necessary. If the user of the Elekta
Neuromag
®
system requires continued
operation during power mains
interruptions, it is recommended that the
Elekta Neuromag
®
system be powered
from an uninterruptible power supply.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Not applicable
Not applicable
The system is intended to measure
magnetic fields at a very low level in a
frequency range that includes 50/60 Hz.
A magnetic site survey and adjacent
determination of necessary magnetic
shielding must be performed for each
installation site prior to installation as a
normal part of the site planning.