28
NM20216A-F
1
Technical data
Guidance and manufacturer's declaration – electromagnetic immunity
The Elekta Neuromag
®
system is intended for use in electromagnetic environment specified below. The
customer or the user of the Elekta Neuromag
®
system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Elekta Neuromag
®
system
combined with the magnetically shielded
room (see 1.14.1), including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
d
= 1.2
√
P
Radiated RF
IEC 61000-4-6
3 V/m
80 MHz to 2.5 GHz
3 V/m
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 2.3
√
P
800 MHz to 2.5 GHz
where
P
is the maximum output power rating
of the transmitter in Watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
[Elekta Neuromag
®
system is used exceeds the applicable RF compliance level above, the Elekta
Neuromag
®
system should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the Elekta Neuromag
®
system
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.