Technical data
1.14.3. Guidance and manufacturer’s declaration
NOTE!
The Elekta Neuromag MEG system should not be used adjacent
to other systems, and if adjacent operation is necessary, the system should
be observed to verify normal operation in the configuration in which it
will be used.
NOTE!
The Elekta Neuromag
®
system should only be used permanently
installed together with magnetically shielded room, see 1.14.1. It is
essential that the actual RF shielding effectiveness and filter attenuation
of the shielded location be verified to ensure that they meet or exceed the
specified minimum values.
NOTE!
Inside the magnetically shielded room, the use of devices
radiating RF energy such as portable and mobile RF communications
equipment such as mobile phones is prohibited. Use of other equipment
except accessories and options supplied with the system should be
avoided; only equipment verified not to affect system performance may
be used. A notice concerning this should be posted at the entrance of the
magnetically shielded room.
Guidance and manufacturer's declaration – electromagnetic emissions
The Elekta Neuromag
®
system is intended for use in electromagnetic environment specified below.
the customer or the user of the Elekta Neuromag
®
system should assure that it is used in such an
environment.
Emissions tes
t
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Elekta Neuromag
®
system uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The Elekta Neuromag
®
system, when installed properly as
described in 1.14.1 is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage
fluctuations / flicker
emissions
IEC61000-3-3
Not applicable