22
NM20216A-F
1
Technical data
• Backup device: compressing DAT tape drive (option)
• CD-ROM Drive: 650 MB
• Expandability: additional workstations
• Magnetic resonance image (MRI) importing: DICOM 3.0 supported
(Conformance Statement available upon request)
• Peripherals (optional): full range available
• Hardcopy devices (optional): black-and-white laser printers, color
solid-ink printers
• Data transfer to other systems
• Ethernet connection to data acquisition system via a media converter
(optical LAN link)
1.9.
Helium transfer equipment
• From standard storage Dewar with flexible siphon, connecting to
integral fixed siphon part in the probe unit
• Flexible siphon
• 2 filter cartridges
• Flexible siphon extension tube
• Cryogenics kit
• Transfer exhaust hose (silicon) with hose clamp
• Pressurizing unit
1.10. Environmental and power requirements, grounding
• Temperature during operation (performance guaranteed):
+20
°
C…+24
°
C
• Temperature during operation (performance not guaranteed, IEC
60601-1 requirements fulfilled): +10
°
C…+40
°
C
• Temperature during storage and transport: +0
°
C…+40
°
C
• Relative humidity during operation (performance guaranteed):
40%…70% RH, non-condensing
• Relative humidity during operation (performance not guaranteed, IEC
60601-1 requirements fulfilled): 30%…75% RH, non-condensing
• Relative humidity during storage and transport: 10%…95% RH, non-
condensing. Special packaging instructions must be obeyed (available
from Elekta Neuromag Oy).
• Mains power voltage: 100/115/200/230/240 V~
±
10 %
• Mains frequency: 47…63 Hz