User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
51
Handheld mode measurement
4.2 ABR
A second click on the “Latency” button enables to dis-
play a latency graph for the I, III and V waves.
The "
Delta
" option displays a table summarizing the
Delta results between the left and right measurement
waves (at the same intensity).
The red boxes indicate any Delta figures above "nor-
mal" (set at 0,3
ms
).
Gray boxes with the letters "N/C" indicate that there is
not enough information to calculate the corresponding
Delta (there are missing markers).
Finally, the "
Curve
" option is used for selecting the curves one by one in order to display them on a full-screen. This
gives a more detailed curve, for instance, to place the markers more accurately. Click on return to come back to the multi-
curve window.
When several iterations of the same power are selected, they appear on the same graph. The placement of
markers and the calculation of deltas are done on the first curve of the graph (red or blue curve).