Report Sensor Health Status (version 3.15 and higher)
UuuH?
uu = two-digit unit number
Response:
grp0=ttt-bbb--grp1=ttt-bbb
grp3=ttt-bbb--grp4=ttt-bbb
UuuDlll.ll[Dlll.ll]Fttt[Fttt……]EeeeeWwwww
BATTERY-VOLTAGE:vv.vV
If sensor works in normal operation parameters the message will be
UuuSENSOR-OK
Error, one or more of the following error messages
NO-SWITCH-CLOSED
ONLY-ONE-GROUP-FOUND-ON-A-TWO-FLOAT-SYSTEM
TOO-MANY-GROUPS-TO-RESOLVE-THE-LEVEL
NO-FLOAT-CONFIGURED
TRANSMIT-TO-PIC-PROCESSOR-FOR-LEVEL-FAILED
RECEIVE-FROM-PIC-PROCESSOR-FOR-LEVEL-FAILED
TRANSMIT-TO-PIC-PROCESSOR-FOR-TEMP-FAILED
RECEIVE-FROM-PIC-PROCESSOR-FOR-TEMP-FAILED
Warning, one or all of the following warning messages:
WARNING!-POSSIBLE-LEVEL-DEGRADATION
WARNING!-POSSIBLE-LEVEL-DEGRADATION-DUE-TO-OFFSET
Report the Level Error Setting (Version 3.09 and Higher)
UuuSETERR?
Response: UuuSETERR=x
uu = unit number
x = 0 is set for level error report to be 999.99 (default)
x = 1 is set for level error report to be 000.00
Report the Modbus 16 bit Unsigned Integer, 32 bit or 2 x 16 bit Floating Point Mode
UuuIF?
Response: UuuIF=x
uu = unit number
x = 0 is set for 16 bit Unsigned integer mode. This is the default setting.
x = 1 is set for 32 bit floating point mode.
x = 2 is set for 2x16 bit floating point mode (v3.14 and higher)
Report Battery Voltage
UuuBV?
Response: UuuBVvv.vV
uu = unit number
vv.v = battery voltage in volts