DIAMAG
Operating Manual
Page 19
4.2. Device electric power consumption: 10 VA.
4.3. The device generates pulsed magnetic fields of the following
types and parameters:
a) field types:
-
«travelling», when there is a serial actuation of all the emitters of
the flexible emitting lines from the 1
st
emitter to the 6
th
in cycles (pro-
grams No.1, No.2, No.3);
-
«static», when there is a simultaneous actuation of all the emit-
ters (program No.4).
b) the frequency of emitter excitation with single pulses:
- 7 pulses / sec (program No.2)
Relative deviation of the frequency rate is within
5%.
c) the frequency of emitter excitation with pulse packets changes au-
tomatically as follows:
- from 1 to 5 pulses / sec (programs No.1, No.4);
- from 5 to 15 pulses / sec (program No.3);
Relative deviation of the frequency rate extremes is within
10%.
d) the time during which the repetition frequency of pulse packets
changes from maximal to minimal value and back is: 126
10 sec for
program No.3 and 612
30 sec for programs 1 and 4;
e) pulse frequency within each packet:
- 30 pulses / sec (program No.3);
- 7 pulses / sec (programs No.1, No.4);
Relative deviation of the frequency value is within
5%;
f) peak values of the magnetic field density
on the emitters’ surface
are:
- 10 mT (programs No.1, No.2, No.3);
- 8 mT (program No.4);
Absolute deviation of the field density peak value (B) is within
[0.2B+0.6] mT.
4.4. The device provides storage of 4 preset exposure programs in
its non-volatile memory.
4.5. The non-volatile memory of the device allows storage and dis-
play (after plugging-in) of the last used program number.