DIAMAG
Operating Manual
Page 18
way that the No. 1 emitters are located over the occipital region, with
the
«N» (north) side facing the head. The methodology is the same.
The duration of exposure of the spine and head during the first 3 treat-
ments is 10 minutes; the subsequent procedures are to last 15 minutes
for each region.
The procedures are carried out once a day. The course length is 10-
20 days.
MAINTENANCE SUPPORT
Maintenance of the DIAMAG device includes preventive inspection,
cleaning from dust and dirt, disinfection, and periodic control of its op-
erability.
The periodic control of the device operability is to be carried out at
least once a year. For this purpose, do the following:
- arrange the flexible emitting lines in a way to provide access to all
the emitters;
- plug the device into the mains;
- select one of the four exposure programs;
- activate the magneto-action;
- using the magnetic field indicator, check the presence of magnetic
field in each of the emitters;
- stop the action;
- unplug the device from the mains.
A preventive (maintenance) inspection is to be performed at least
once every three months. During the inspection, it is necessary to
check the integrity of the cables, the plug, the mains cord, the control
unit housing, and of the emitters of the flexible emitting lines. Disinfec-
tion is made as may be necessary.
SPECIFICATIONS
4.1. The device is functional with power supply from alternating
current mains of:
~220V (-10%; +10%) or ~230V (-10%; +6%), frequency 50Hz
~120V, frequency 60Hz